Hotplates & Stirrers
Introduction
Thank you for purchasing this Cole-Parmer product. To get the best performance from the
equipment, and for your own safety, please read these instructions carefully before use.
Before discarding the packaging check that all parts are present and correct.
This equipment is designed to operate under the following conditions:
❖
For indoor use only
❖
Use in a well ventilated area
❖
Ambient temperature range +5°C to +40°C (+41°F to 104°F)
❖
Altitude to 2000m, (6500ft)
❖
Relative humidity not exceeding 80%
❖
Power supply fluctuations not exceeding 10% of nominal
❖
Overvoltage category II IEC60364-4-443
❖
Pollution degree 2 IEC664
❖
Use with a minimum distance all round of 200mm, (8 in.) from walls or other items
If the equipment is not used in the manner described in this manual and with accessories other than
those recommended by the manufacturer, the protection provided may be impaired.
Electrical Installation
THIS EQUIPMENT MUST BE GROUNDED
Before connection please ensure that the line supply corresponds to that shown on the
rating plate located on the base of the unit.
Power requirements:
230v Models
120v Models
04801-22
500W
04801-20
500W
04801-26
700W
04801-24
700W
04801-52
50W
04801-50
50W
04801-56
50W
04801-54
50W
04801-62
550W
04801-60
550W
04801-67
750W
04801-65
750W
04801-66
550W
04801-64 550W
04801-70
750W
04801-68 750W
There is an IEC hot condition power socket (C16) for a hot condition IEC plug (C15).
The 230V unit is provided with a power/mains lead fitted with an IEC hot condition (C15) plug and
with either a UK 3-pin plug and a “Shuko” 2-pin plug for connection to the power supply.
The 120V unit is provided with a power/mains lead fitted with an IEC hot condition (C15) plug and
with a NEMA 5-15 plug for connection to the power supply.
Should the cable not be suitable for connecting to the power supply, replace the plug with a suitable
alternative.
THIS OPERATION SHOULD ONLY BE UNDERTAKEN BY A QUALIFIED ELECTRICIAN.
3