Chapter 6
System Commands
© 2010 Colby Instruments. PDL-100A Operating and Programming Manual Page 49
INC
INCREMENT
COMMAND
Increment the delay setting by step size amount.
Command
Syntax
INC
Example
INC
Returns
None
Programming
Example
HP : OUTPUT 705;"INC"<NL>
NI : CWGPIB.Write “INC”
MODE
MODE
COMMAND
Set the operating mode for the PDL-100A.
SERIAL mode step size increment is 0.50 ps
PARALLEL mode step size increment is 0.25 ps and
total delay range is 0 to 312.50 ps.
Command
Syntax
MODE [SER | PAR | 312.5ps | 625ps]
SER Set the PDL-100A to Serial Trombone Mode
PAR Set the PDL-100A to Parallel Trombone Mode
312.5ps Set the PDL-100A to Parallel Trombone Mode
625ps Set the PDL-100A to Serial Trombone Mode
Example
MODE 312.50ps
MODE SER
MODE 625PS
Returns
None
Programming
Example
HP : OUTPUT 705;"MODE SER"<NL>
NI : CWGPIB.Write “MODE PAR”