PXIe Platform Bit Error Rate Tester
|
BERTPXIe-1003/1005
Coherent Solutions Ltd.
Version
1.00
88
Example: Control of the BERTPXIe modules with SCPI
The examples in this section show how you can control the BERTPXIe modules using SCPI
commands.
To debug unresponsive or incorrect commands that have been sent to the module, you are
recommended to use the
*ESR?
command to check the standard event status after each
command.
Using slot 10 as an example, take the following steps to make sure that the instrument is
ready:
1.
Check that all modules are ready to run commands:
*OPC?
2.
Check the operational status of the module installed on slot 10:
:SLOT10:OPC?
3.
Check the standard event status register to make sure the commands have been run
with no errors:
*ESR?
Take the following steps to check the error detection measurements, assuming slot 10 and
channel 2 are used. Before running these commands, make sure the instrument is ready.
1.
Start error detection:
:SENS10:MEAS2:STARt
2.
Check the total error count:
:CALCulate10:DATA2:EALARM? ERROR
3.
Check the total number of bits transferred:
:CALCulate10:DATA2:EALARM? BITS
4.
Check the bit error rate:
:CALCulate10:DATA2:EALARM? BER
5.
Check all the error-detection measurement information:
:CALCulate10:DATA2:EALARM? FULL
6.
Stop error detection:
:SENSe10:MEASure2:STOP
Take the following steps to acquire eye scan measurement, assuming slot 10 and channel 2 is
used. Before running these commands, make sure the instrument is ready.
1.
Set the vertical resolution step size of the eye measurement to the maximum value:
:SENSe10:MEASure2:EDIagram:VERTical MAX