![Coherent Solutions BERTPXIe-1003 Скачать руководство пользователя страница 79](http://html1.mh-extra.com/html/coherent-solutions/bertpxie-1003/bertpxie-1003_user-manual_2634920079.webp)
PXIe Platform Bit Error Rate Tester
|
BERTPXIe-1003/1005
Coherent Solutions Ltd.
Version
1.00
78
9.4.39
:OUTPut<n>:CLOCk:POWer
Command :OUTPut<n>:CLOCk:POWer
Description Sets the power for the RF output clock.
Syntax
:OUTPut<n>:CLOCk:POWer<wsp><value|MIN|MAX|DEF>
Parameters
•
value: The desired new set point
•
MIN: Sets to the minimum value
•
MAX: Sets to the maximum value
•
DEF: Restores the default value
Response N/A
Example
:OUTPut<n>:CLOCk:POWer 0.0
9.4.40
:OUTPut<n>:DATA<m>:AMPlitude?
Command :OUTPut<n>:DATA<m>:AMPlitude?
Description Queries the voltage amplitude of the output signal.
Syntax
:OUTPut<n>:DATA<m>:AMPlitude?
[<wsp><MIN|MAX|DEF|SET|ALL|LIST|STEP|UNIT>]
Parameters
•
MIN: Queries the minimum configurable value
•
MAX: Queries the maximum configurable value
•
DEF: Queries the default power-on value
•
SET: Queries the current SET value
•
LIST: Queries the list of discrete options for configurable values (if
available)
•
ALL: Queries the values specified by MIN, MAX, DEF, SET, and LIST
•
STEP: Queries the minimum resolution of the amplitude value
•
UNIT: Queries the default unit for the amplitude value
By default, this command queries the current SET value.
Response The response is a single value, or a comma separated array. When
ALL
is requested, the format is "
<MIN>,<MAX>,<DEF>,<SET>,<LIST>
"
Example
:OUTPut3:DATA2:AMPlitude? ALL
->200,1100,200,200,NAN
NOTE: This does not have a discrete LIST.