Connecting to an IV Pole
1. Rotate the ratchet handle on the clamp to open up the
jaws. Connect the clamp mount assembly to the IV pole
and secure the clamp tightly by turning the ratchet
handle.
2. Next, connect the back mounting plate to the scanner by
pressing and sliding down the tabs into the back of the
SLS-WAVE. Make sure that the USB cable is routed over
the bracket as shown below.
3. Insert the screw through the mounting bracket and
tighten to the SLS-WAVE scanner with the L-key.
4. Next, remove the protective film liner from the SLS-
WAVE. Attach the SLS-WAVE assembly to the clamp
mount assembly, already attached to the IV pole, by
rotating the platform thumbwheel as shown below.
5. The SLS-WAVE is now mounted to the IV pole.
6. Insert the SLS-WAVE USB into the AIMS computer.
Configuring SLS-WAVE with Cerner
1. Work with your Cerner representative to ensure
SurgiNet system is configured for scanning and that the
minimum software installed is January 2018 or newer.
2. SLS-WAVE is delivered configured to support scanning
of 1-D barcodes. If at some point the SLS-WAVE needs to
be configured again, then scan the following barcode.
WARNING: The scanner should be pointing down and
positioned to keep the scanner light from shining in the
patient’s or user’s eyes.
Note: USB extension cords are available if required to
reach from the SLS-WAVE to the AIMS computer. They
can be ordered using Codonics Part Number SLS-
WAVE-USB-EXT.