Instruction manual of distancemeter TEPEX 2S - May 2021 version
Page 12 / 12
I APPENDIX
HOW TO CHECK THAT THE TEPEX 2S IS WORKING PROPERLY
The distance
D
between the probe and the reading box must be 8.00 metres. The
probe must be vertical as well as the reading box. They must be positioned at the same
height.
Put new batteries in the reading box and in the probe.
The test must be carried out on natural horizontal ground free of magnetic or
electromagnetic disturbance.
HNI
Ferrous slab, soil containing ferrite,
presence of bulky metal objects...