Care Essentials
CWS4000 Cocoon Warming System Operator Manual
QPF 053 Revision S
Date 07.05.2012 Supersedes: 08.02.2012
Page 13
8.3.
Commingling Recommended Instructions
Cocoon CWS4000 Warming Unit
Brand of Blanket
Recommended Instructions
Tyco
WarmTouch™ Blankets
Insert the CWS4000 delivery hose nozzle well into the
WarmTouch blanket.
Secure the blanket to the nozzle with adhesive tape.
Inflate the blanket and ensure that there is no air leakage.
Refer to Figure 6 below.
Arizant
Bair Hugger™ Blankets
Ensure that the Care Essentials CWS203 adaptor is attached
to the CWS4000 delivery hose nozzle.
Insert the adaptor into the connection point of the Bair Hugger
blanket. The adaptor should fit securely and be held in place
by the blanket connection.
Inflate the blanket and ensure that there is no air leakage.
Refer to Figure 7 below.
Figure 6 Cocoon CWS4000 warming unit
attached to WarmTouch™ blanket
Figure 7 Cocoon CWS4000 warming unit
attached to Bair Hugger™ blanket