Calibration M ode
CAUTION: TO CALI BRATE THE LIFEGUARD™ YOU MUST USE
100 % OXY GEN.
The LIFEGUARD™ uses a two-point calibration. The First point is and
must be
21% while the Second point is and
must be
100 %. Calibrate the
LIFEGUARD™ as follows:
1. Attach sensor and cable to your LIFEGUARD™. If you are calibrating a new
O
2
sensor allow the sensor to stabilize in air for approximately 24 hrs after
removal from the package.
2. Have a source of
pure (100%) O
2
at hand for use in Step 6, below.
3. Expose the sensor to
atmospheric air (21%)
and allow its readings to
stabilize (see Figure 20a).
Fig-20a
Calibration Mode (Sensor exposed to 21% - AIR)
4. Attach the Field Programmer probe to the CTU and verify the WU display
changes to show “
C-2
” in the upper left corner (see figure 18, page 29) to
indicate start of PROGRAMMING/CALIBRATION Mode. If not, tap the PWR
circle a single time and confirm PROGRAMMING/CALIBRATION Mode (by
noting presence of the “
C-2
”).
5. Press the SELECT button as described in the Programming Section,
stepping through each of the following:
•
C-2
•
C-1
•
A-3
•
A-2