Low and High Oxygen Level Alarm 

The SAN-102 is equipped with audible and visual alarms to alert you when the ambient oxygen 
concentration exceeds either of the two factory preset alarm levels:  

Danger Low Oxygen Alarm:

 19.5% - LED will flash and audible alarm will sound 3x / sec.

Warning High Oxygen Alarm:

 23.5% - LED will flash and audible alarm will sound 2x / sec.

Normal atmosphere contains between 20.8% and 21% oxygen. OSHA defines any atmosphere 
that contains less than 19.5% oxygen as “oxygen deficient” and any atmosphere that contains 
more than 23.5 percent oxygen as “oxygen enriched.” These levels are factory set and cannot 
be changed.  

Fahrenheit and Celsius Switching 

Press the °F/°C button to switch between the two temperature units.

Press the Power button briefly to save the temperature setting and return to normal display. 

Low Battery Alarm 

The meter continuously measures the rechargeable battery’s voltage. If it drops to a critical 
level, the low battery alarm will sound until the battery is recharged or the battery power is 
depleted.  

Note that the meter can continue to monitor oxygen levels while charging. 

CLEANING AND STORAGE 

The front panel and case can be cleaned with a mild solution of detergent and water. Apply 
sparingly with a soft cloth and allow drying completely before using. Do not use aromatic 
hydrocarbons or chlorinated solvents for cleaning. 

ATTENTION! - If a Warning or Danger alarm is triggered 

while using the instrument as a safety monitor, leave 

the area and seek fresh air immediately. Remaining on 

site under such circumstances can cause serious 

impairment or even lead to death. 

Calibration

The SAN-102 comes pre-calibrated from the factory

.

 However, the O2 sensor should be 

calibrated at least once a year, or as described in your company's safety procedures. You can 

perform the calibration yourself, or you can return it to CO2Meter for factory calibration at a 

nominal fee.
SAN-102 can be atmospherically calibrated. By pressing the 

Fahrenheit and Celsius 

switching button

 for 4 seconds, the monitor will enter O2 atmospherically calibration mode. 

While in calibration mode, press the 

Fahrenheit and Celsius switching button

 quickly to 

switch between atmospherically calibration or Exit. Press the 

Power button

 to confirm.

 

The atmospherically calibration will not require additional materials but must be performed in 

outdoor (fresh air) during the entire procedure.

 

The temperature and humidity sensors do not require calibration and should remain accurate 

for the life of your unit.