16
– Default Syringes
T34L™
Operator Manual
27
16
– Default Syringes
Default Syringe Brands Configured for Use with
T34L
TM
Braun Omnifix 2, 5, 10, 20, 30 & 50ml
BD Plastipak 3, 5,10, 20, 30ml & 50ml
Monoject 6, 12, 20, 35 & 50ml
Codan/Once 2.5, 5, 10, 20 & 50ml
Terumo 5, 10, 20, 30 & 50ml
It is possible to disable default syringes from the memory or replace them with one not listed above. This
procedure is not detailed in this manual as it should only be undertaken by trained, certified service
centers or biomedical engineers. Please consult a biomed engineer or your local
T34L
TM
distributor should
this need arise.
Time to Alarm from Occlusion
The Time to Alarm from Occlusion for the T34L Syringe Pump are in accordance to the table below.
All the tests were performed using a CME extension set, 100-172S and a BD Plastipak syringes filled with
water.
Time to Alarm from Occlusion
Manufacturer
Size
Rate
300mmHg
600mmHg
900mmHg
BD Plastipak
5ml
1ml/h
11'
11'50"
14'30"
100ml/h
7"
10"
10"
10ml
1ml/h
17'30"
20'25"
31'30"
100ml/h
8"
11"
13"
20ml
1ml/h
29'30"
42'30"
58'45"
100ml/h
19"
20"
23"
50/60ml
1ml/h
46'
54'30"
117'
100ml/h
28"
34"
36"
The tolerance for the BD Plastipak 50/60mll syringe is ±25% or ± 100mmHg
The tolerance of the BD Plastipak 5ml syringe is ±25% ±200mmHg