– Infusion Set Up & Programming
T34L™
Operator Manual
17
Duration Changed
Please Check rate
Volume 12 ml
Duration 24:00
Rate 0.5 ml/h
Confirm,
Press YES
Start Infusion?
Time Remaining 23:57
Rate 0.5ml/hr
<<<<< Pump Delivering
Time Remaining 23:58
Rate 0.5ml/hr
10ml BD Plastipak
NOTE:
In
VOLUME OVER TIME
configuration (UNLOCKED)
the user can only set the rate by selecting the required duration.
If a change of rate is required, press NO to back up to the
duration screen and amend duration as required. Upon
confirmation of duration change, the pump alerts the user (right)
to check the new rate which is then displayed.
NOTE:
In
VOLUME OVER TIME
Senior clinical personnel or technicians may have set a
MAXIMUM RATE LIMIT to prevent
T34L
TM
being programmed at rates above a fixed limit in line
with the application the pump is regularly used for. Pump will not allow the user to scroll to a
duration that requires a rate above this limit to deliver the contents of the syringe over that time.
19. The Summary Screen displays the Volume (To Be Infused), Duration
and Infusion rate.
Warning:
Always
take time to check the contents of the summary screen match the
prescription before starting infusion.
20. Press YES/START to confirm acceptance of the pre-set (if Program Lock ON) or programmed (if LOCK
OFF) infusion parameters.
21. Pump prompts START INFUSION? Check administration set is
attached to patient access device and the clamp is released. Press
YES/START to commence infusion
22. Whilst infusing the display shows Infusion Time Remaining (top line), Infusion Rate (main/central line)
and the bottom line will alternate between Syringe Size & Brand and a delivery animation to indicate the
pump is running (in case the LED indicator on the keypad has been disabled to conserve battery life (see
Change Set Up).