Change the reference temperature for the temperature compensation algorithm
Note:
This feature is only available in communication firmware version > 3.0!
- Change the reference temperature:
o
Type in the new reference temperature
o
Click
on Button “Set. Ref. Temp.” to set the new reference temperature,
the communication unit will be restarted and an automatic T-Reset is
performed at the same time. So the temperature algorithm need new
learning data (i.e. performing a measurement as noted on the previous
page).
Adjust the email notification settings
- Fill in your email settings to enable automatic email notification
-
Alarm Report notification is enabled by click on Button “ E-Mail Alarm Report”
- Cyclic automatic email notification is enabled
by click on Button “ Automatic E-
Mail Report”
-
The interval of email notifications can be modified by the field “E-Mail interval
[min]”. The unit is minutes.
Adjust the limit settings of the temperature compensated data
The limit values of the temperature compensated values of
TC
,
rTC
, T, gradients of
TC
&
rTC
can be set individually. Please note that this feature will be available after
the temperature compensation has been initialized properly as described on the
previous page. These limit values can vary on the different oil types used in the
monitored system and have to be adjusted on the specific application.
Change the language
Two languages are supported: German, English.
-
Choose the language by click on the Button “Language”.
Close the software
Exit the software by click on the button
“Exit”. A window will open: “Overwrite limit
settings?” You can choose “Yes!” or “No!” to save the last used settings to the
Computer. At the next program start the last saved settings are reloaded.