Emissions Guidance Table 

Gui

dance and manufa

cturer’s declaration 

-

 Electromagnetic emissions 

 

The Ceevac suction pump is intended for use in the electromagnetic environment specified 

below. The customer or user of the Ceevac suction pump should assure that it is used in such 

an environment.

 

Emissions Test

 

Compliance

 

Electromagnetic environment 

-

 guidance

 

Irradiated / Conducted 

Emissions

 

CISPR11

 

Group 1

 

The Ceevac suction pump uses RF energy only 

for its internal functioning. Its RF emissions are 

very low and will not cause interference in 

proximity of any electronic appliance.

 

Irradiated / Conducted 

Emissions

 

CISPR11

 

Class B

 

Harmonic Emissions

 

IEC 61000

-

3

-

2

 

Class A

 

Voltage fluctuations / flicker 

emissions

 

IEC 61000

-

3

-

3

 

Complies

 

The Ceevac suction pump is suitable for use in 

all establishments, including domestic 

establishments and those directly connected to 

the public low

-

voltage power supply network that 

supplies buildings used for domestic purposes.  

 

EMC Information Tables per EN 60601

-

1

-

2:2014.

 

 

In accordance with EN 60601

-

1

-

2:2014 Medical electrical equipment 

-

 Part 1

-

2: General 

requirements for safety 

-

 Collateral standard: Electromagnetic compatibility 

-

 Requirements and 

tests

 

1) “Medical Electrical Equipment needs special precautions regarding EMC and needs to be 

installed and put into service according to the EMC information provided in the Accompanying 

Documents” (the following tables).

 

2) “Portable and Mobile RF Communications Equipment can affect Medical Electrical 

Equipment.”

 

3) “The equipment or system should not be used adjacent to or stacked with other equipment 

and that if adjacent or stacked use is necessary, the equipment or system should be observed to 

verify normal operation in the configuration in which it is used.”

 

The following tables provide information regarding the EMC characteristics of this Medical 

Electrical Equipment

.

 

The performance of all functions of the Ceevac suction pump  are considered essential 

performance for the purpose of electromagnetic immunity.

 

Содержание SUC 81030

Страница 1: ...Ceevac High Vacuum High Flow Portable Suction Pump SUC 81030 12 VDC 100 240 V 50 60 Hz User Manual CLEMENTS Manual No SUC 81030 008 Issue 2...

Страница 2: ...by authorised personnel in accordance with the instructions as described in this manual Operated in this way the Ceevac High Suction Pump will provide the standard of service specified Due to continu...

Страница 3: ...and Operation 10 Using FLOVAC Disposable Liners 11 Battery Status and Charging 13 Cleaning 15 Spare Parts 17 Troubleshooting 18 Wiring Diagram 19 Diagnostic Setup 20 Return for Repair 20 Periodic Safe...

Страница 4: ...ly trained and authorized personnel Contraindications Before using the Ceevac consult the instructions for use Failure to follow instructions in this manual could cause harm Do not use Ceevac for thor...

Страница 5: ...rning time Minimum 5 minutes Power Pack Battery Charge External Powers pump Monitors and charges battery when mains power supply is connected Pump Piston type Filter Disposable hydrophobic and bacteri...

Страница 6: ...g table In addition for vehicular transportation the unit should be kept upright State Parameter Minimum Maximum Operating Temperature 5 C 35 C Humidity 10 RH 93 RH Barometric Pressure 700 hPa 1060 hP...

Страница 7: ...tion cannula Insulation Class II double insulation Hz Mains power frequency Direct current Battery On Off CE Mark in conformity with CE MEDDEV directive 93 42 CEE and subsequent changes Manufacturer R...

Страница 8: ...rk Degree of protection against intrusion by body parts or objects an against ingress by water 1st Digit Penetration of Solids 2 Protected against solids 12 5 mm 2nd Digit Penetration of Liquids 1 Pro...

Страница 9: ...ack charger 2 Confirm that electrical rating of power pack matches that of the mains power Avoid the use of power boards and extensions 3 If power pack charger is used maintain clear access to the mai...

Страница 10: ...let 3 Connect the long silicone tubing to the jar PATIENT port 4 Select power source 1 Connect plug of supplied power pack charger to power socket at rear of pump OR 2 Connect plug of supplied cigaret...

Страница 11: ...iners 1 Remove liner from package and extend liner from lid 2 Insert liner bag 1 into the matching size support jar 2 Press edge of lid firmly all around edge to ensure seal 3 Close the TANDEM inlet 3...

Страница 12: ...e may be signs of decay and replacement is recommended Silicone Tubing The number of cleaning and sterilization cycles of the silicone tubing is dependent on the actual usage Tubing should be checked...

Страница 13: ...battery charge Battery Charging Use only the supplied power pack charger model UE60 140429SPA1 Recharge the battery after each use to maintain the pump s maximum standby capacity When the battery is...

Страница 14: ...ts Battery Service Life The following chart shows the expected service life of the battery against the depth of discharge In summary maximum service life of the battery will be obtained from a usage p...

Страница 15: ...rilised with sterilants determined by local protocols and compatible with silicone tubing Suction tubing may be autoclaved to a maximum of 121 C for 15 minutes At higher temperatures the suction tubin...

Страница 16: ...scoloured Also check the tubing nipples Check the seal ring on the lid and replace if hard cracked or perished Check the fit of the lid in the collection jar canister Check the overflow cut off valve...

Страница 17: ...er Pack 1 SUC 81030 014 PCBA Electronic Control for Ceevac SUC 81030 015 Ba8ery SLA 12V 4 0Ah for Ceevac SUC 81030 016 Adapter Vehicle 12V Fused for Ceevac SUC 81030 017 Switch ON OFF for Ceevac SUC 8...

Страница 18: ...the Control Panel indicates zero If not release vacuum and switch pump off and on 2 Pump Running But No Vacuum Check that all fittings are connected tightly Turn the vacuum controller knob on the cont...

Страница 19: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 19 Wiring Diagram Wiring Diagram...

Страница 20: ...Disconnect all items from inlet and with pump switched off confirm that gauge reads zero A non zero reading indicates a faulty gauge Set vacuum control knob to maximum switch on pump and confirm that...

Страница 21: ...indicates leaks in jar or connections 1 Occluded leak check Flow Check Max Vacuum Gauge Connect pump as shown in diagram Set vacuum control knob to maximum and switch on pump Connect jar inlet to cali...

Страница 22: ...eturn for Repair Before returning the ACeevac for repair the external surfaces and any accessories must be carefully disinfected with a cloth soaked in methylated spirits or hypochlorite based solutio...

Страница 23: ...voltage power supply network that supplies buildings used for domestic purposes EMC Information Tables per EN 60601 1 2 2014 In accordance with EN 60601 1 2 2014 Medical electrical equipment Part 1 2...

Страница 24: ...put output lines The device doesn t change its state Mains power quality should be that of a typical commercial environment or hospital Surge EN 61000 4 5 1 kV differential mode The device doesn t cha...

Страница 25: ...the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol Radiated Immunity EN 61000 4 3 10 V m 80 MHz to 2 7 GHz for non life s...

Страница 26: ...ansmitter W 150 kHz to 80 MHz d 3 5 V1 P 80 MHz to 800 MHz d 12 E1 P 800 MHz to 2 7 GHz d 23 E1 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters with...

Страница 27: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 27 Notes Notes...

Страница 28: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 28 Notes Notes...

Страница 29: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 29 Notes Notes...

Страница 30: ...ble to the original purchaser of the product 3 The product must not have had its serial number removed defaced or changed its casing opened its power supply altered or have been tampered with in any o...

Страница 31: ...rms of the sale of the product or otherwise in connection with the product is limited to the amount paid by the purchaser to ICU Medical Australia for the product 10 Apart from any liability imposed b...

Страница 32: ......

Отзывы: