Suction Accessories
5.
Press switch to turn on pump. (Press again to turn off.)
6.
Occlude tubing and adjust vacuum control to the required
vacuum level. (Clockwise = increase)
Always reduce the vacuum level for paediatric and neonatal
patients.
7.
Apply suction to patient with compatible suction cannula.
8.
When finished, switch off pump by pressing switch.
Ensure that the mains power plug is accessible at all times when
pump is in use, in case a positive disconnection from the mains
power supply is required.
Using FLOVAC Disposable Liners
1. Remove liner from package and
extend liner from lid.
2. Insert liner bag [1] into the matching
size support jar [2].
Press edge of lid firmly all around
edge to ensure seal.
3. Close the TANDEM inlet [3] with the
attached cap [4].
4. Fit jar assembly into pump.
5. Press short end of yellow reusable
nipple [6] firmly into matching yellow
port [5] on the lid.
6. Push tubing from filter onto yellow
nipple [6].
7.
Connect patient tubing to white inlet
elbow [7]. Connect patient tubing to
suitable suction cannula.
1
2
7
4
3
5
8
6
Note
: For disposable jars
only
, that have a hydrophobic and
antibacterial filter built-in to the lid, connection may be made directly
to the pump inlet nipple.
For reusable jars, connection
must
be made through an inline
hydrophobic and antibacterial filter.
Содержание SUC 81030
Страница 19: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 19 Wiring Diagram Wiring Diagram...
Страница 27: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 27 Notes Notes...
Страница 28: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 28 Notes Notes...
Страница 29: ...CLEMENTS CLEMENTS CLEMENTS CLEMENTS 29 Notes Notes...
Страница 32: ......