Cleaning and Maintenance

 

Cleaning 

Spills  must  be  cleaned  up  immediately.  The  centrifuge  should  be 
regularly  cleaned  to  maintain  its  appearance  and  to  ensure  that 
dust  or  other  damaging  materials  do  not  build  up  and  affect  the 
performance of the unit. 

• 

Before  cleaning  the  centrifuge  disconnect  it  from  the  mains 
power supply. 

• 

Clean  the  exterior  surfaces  using  a  proprietary  furniture 
polish with a clean soft cloth . 

• 

Wipe the bowl and rotor with a clean soft cloth dampened in a 
solution of warm water and a mild detergent. (If moisture can 
be wrung from the cloth, it is wet, not damp.) 

• 

Most sterilising agents at their standard cleaning dilution can 
be used on the bowl. 

• 

Remove  cleaning  agent  residue  by  wiping  with  a  dry  paper 
towel. 

• 

Solvent-based  cleaning  agents  and  abrasive  cleaning  agents 
should never be used on the Orbital 360. 

 

Maintenance 

The  Orbital  360  is  designed  to  provide  extended  and  reliable 
operation. To ensure proper operation and maximise the life of the 
centrifuge the following requirements must be observed: 

• 

The  centrifuge  should  be  inspected  by  authorised  service 
personnel  every  six  months  as  part  of  your  routine 
maintenance plan 

• 

Confirm  that  lid  interlock  prevents  starting  when  lid  is  un-
locked and also prevents opening lid before rotor has stopped 

• 

Using  an  optical  tachometer  confirm  that  the  speed  remains 
within specification 

• 

Use only genuine Clements spare parts  

Содержание Orbital 360

Страница 1: ...User Manual CLEMENTS Orbital 360 Benchtop Centrifuge CEN 91705 220 240V 50 60Hz Manual No CEN 91705 017 Issue 1...

Страница 2: ...nces St Lidcombe NSW 2141 Australia Phone 61 2 9466 5300 Website www medaust com Safety Thank you for purchasing this Clements Orbital 360 Centrifuge For your safety it is imperative that this unit on...

Страница 3: ...Warnings and Cautions 6 Installation 7 Controls 8 Operation 10 Parameter Setting Menu 12 Operation Summary 14 User Programs 16 Rotors 17 Accessories and Spares 19 Cleaning and Maintenance 20 Fuse Rep...

Страница 4: ...ology for low maintenance operation Accurate digital control of speed and time for repeatable processing High visibility backlit LCD display with touch control panel Electronic lid interlock to ensure...

Страница 5: ...l Protection Class I Noise 65dB A Dimensions 370 W x 250 H x 432 D mm Weight 15kg Environment Indoor 10 to 35 C RH 85 Package Contents Item Description Qty CEN 91705 Cell Cycle 360 Centrifuge 240V 50H...

Страница 6: ...be observed when handling potentially biohazardous materials Carriers with aerosol protection are to be used Fluid Spill Do not allow fluids to enter the centrifuge If a spill occurs wipe clean payin...

Страница 7: ...be placed on a solid level surface in a well ventilated area The back of the centrifuge must be at least 10cm away from any wall or vertical surface Power Requirements Before connecting the centrifug...

Страница 8: ...ss hole Control panel Front View Rear View On Off Switch IEC power inlet Fuse holder Condition LCD Display Shows In Spin Cycle Actual and Set Values of Speed RCF and Time Remaining Setting Mode Set Va...

Страница 9: ...spin cycle PULSE Button Spins rotor when held down ENTER Button Enters and saves changed parameters Exits parameter setting mode Selects user program Exits program mode INC Button Increases value of s...

Страница 10: ...gain being careful not to bend or apply side load to motor shaft 8 Confirm that rotor rotates smoothly and evenly by hand Carrier Fitting swing out style rotors 1 Fit a full complement of compatible c...

Страница 11: ...site positions on the rotor Note Even though the Orbital 360 is designed to absorb considerable out of balance forces running the centrifuge with out of balance samples will cause excessive vibration...

Страница 12: ...itted Note It is essential that the rotor setting matches the rotor fitted Otherwise the rotor could be subject to speeds in excess of its rating leading to damage to the rotor 1 Ensure the rotor is s...

Страница 13: ...ll auto increment or auto decrement if the INC or DEC button is held down 3 Press the ENTER button to save the setting Set Acceleration Profile 1 Press the MENU button until the acceleration field is...

Страница 14: ...d value 2 With no button pressed for 5 seconds unit will time out and revert to previously selected user program Select User Program 1 Press PROG button Press INC or DEC buttons to increment or decrem...

Страница 15: ...or set in the user program selected matches the actual rotor fitted Otherwise the rotor could be subject to speeds in excess of its rating leading to damage to the rotor START Spin As per selected Use...

Страница 16: ...display counting down to zero The rotor will automatically decelerate to zero using the set deceleration profile When the rotor has stopped the beeper will signal the end of the spin cycle 8 Press th...

Страница 17: ...17 CLEMENTS Rotors Angle Rotor CEN 91700 001 24 x 10ml Tubes Max rpm 4000 Max RCF 2850 Rotor Setting 1 Swing Rotor CEN 91700 002 12 x 15ml Carriers Max Speed 5000rpm Max RCF 4100 Rotor Setting 2...

Страница 18: ...18 CLEMENTS Rotors Swing Rotor CEN 91700 004 4 x 50ml Carriers Max Speed 4200rpm Max RCF 2760 Rotor Setting 3 Swing Rotor CEN 91700 003 24 x 5ml Carriers Max Speed 5000rpm Max RCF 4100 Rotor Setting 4...

Страница 19: ...1700 010 Hinge set for Orbital 360 CEN 91700 011 Electronic Lid Lock for Orbital 360 240V CEN 91700 012 Power Transformer for Orbital 260 360 240V CEN 91700 013 Carrier Set 12 x 15ml for Swing Rotor C...

Страница 20: ...rilising agents at their standard cleaning dilution can be used on the bowl Remove cleaning agent residue by wiping with a dry paper towel Solvent based cleaning agents and abrasive cleaning agents sh...

Страница 21: ...tton pressed Close lid Adjust Lid Closed limit switch E6 Overspeed condition Incorrect rotor setting Power control PCB fault Motor fault SWITCH OFF Verify correct rotor setting Replace power PCB Repla...

Страница 22: ...nnectors Rotor does not spin when START button pressed Lid not closed Close lid Loose connecting cables Check and re seat connectors Power transformer fault Replace power transformer Motor fault Repla...

Страница 23: ...23 CLEMENTS 0 1000 2000 3000 4000 0 1000 2000 3000 4000 5000 RCF Graphs 0 20 40 60 80 100 120 140 160 100 200 300 400 500 600 700 800 900 1000 RCF Graphs 5 15 ml 50 ml RCF g rpm 50 ml RCF g rpm...

Страница 24: ...24 CLEMENTS Notes...

Страница 25: ...25 CLEMENTS Notes...

Страница 26: ...rchase Conditions of Warranty 1 The product must be returned to the place of purchase with proof of purchase 2 This warranty is only available to the original purchaser of the product 3 The product mu...

Страница 27: ...lacement of the product within the terms of this warranty and the total liability of BMDi TUTA Healthcare for loss or damage of every kind whether arising pursuant to the terms of the sale of the prod...

Страница 28: ......

Отзывы: