CMS-4 IN-HELMET CARBON MONOXIDE MONITOR
Page 5
© 2019 CLEMCO INDUSTRIES CORP.
www.clemcoindustries.com
Manual No. 29872
1.7.9 Alarm-Lights:
Two LED alarm-lights are located
above the LCD. The
red
alarm-lights show through
the
reflective lenses. The LED’s alert the user to CO
gas, low battery, and failure alarms. The lights flash
once during the startup sequence.
1.7.10 Audible Alarm:
An opening on the left side of
the top case allows the alarm’s sound to exit the case.
The alarm sounds for CO gas, unit malfunctions, low
battery voltage, and as an indicator during normal use of
various display options. It beeps once during the startup
sequence. Refer to Section 3.1 for alarm indications.
1.7.11 Vibrating Alarm:
A vibrating motor mounted
inside the case vibrates for CO gas alarms, unit
malfunctions, and as an indicator during normal use of
the various modes of the CMS-4. It vibrates briefly
during the startup sequence.
1.7.12 Lithium Battery:
A CR2450, 3.0-volt lithium
coin battery powers the CMS-4. The battery icon in the
upper right of the LCD shows remaining battery life.
When the CMS-4 detects a low battery voltage, a low
battery warning is activated. When battery power is too
low for Measuring Mode, the CMS-4 sounds a dead
battery alarm. (Gas reading disappears,
FAIL
appears in
the middle of the screen,
BATTERY
appears at the
bottom of the screen, LEDs flash once per second, and
audible alarm double pulses once per second.)
A spare battery is provided with the monitor;
refer to
Section 6.1 for instructions on replacing the battery.
2.0
SETUP ‒ INSTALLATION, STARTUP AND
OPERATION
2.1
Installation
This section explains the initial installation of the CMS-4
to the inside of the respirator.
2.1.1
Prepare Mounting Clip
2.1.1.1
Adhere Velcro
®
loop to CMS-4 mounting clip:
Trim the Velcro
®
loop strip to 3/4" x 1". Remove the
backing and adhere it to the bottom of the monitors
mounting clip, as shown in Figure 4.
2.1.2 Installing Monitor in Apollo 20 Respirator
Refer to Section 2.1.3 for installing the monitor in Apollo 60
and Apollo 600 respirators.
2.1.2.1
If not already done, adhere the Velcro
®
loop to
CMS-4 mounting clip, per Section 2.1.1.
Figure 4
2.1.2.2
Open the lens frame and remove the lenses.
2.1.2.3
Don (put on) the helmet and temporarily place the
monitor on one of the side walls of the window extension,
as shown in Figure 5. Make sure the monitor’s LED lights
are visible.
2.1.2.4
Mark the location and remove the monitor and helmet.
Figure 5
2.1.2.5
Remove backing from the Velcro
®
hook tape and
adhere it at the marked location (the loop tape adheres to
the mounting clip on the monitor). Use the Velcro
®
to
attach the monitor to the hood, positioning it so the
control buttons face toward lenses and alarm-lights
toward the inside of the respirator. Don the helmet to
make sure the monitor’s alarm-lights are visible from
inside the helmet.
2.1.2.6
Remove the monitor and prepare it for operation,
per Section 2.2.
2.1.2.7
Replace the lenses and lens frame assembly.
2.1.2.8
After the initial setup is done, remove and
reattach the monitor using the Velcro
®
.
Remove the
Make sure the monitor’s
LED lights are visible when
the helmet is donned, mark
its location inside of the
window extension, and
attach Velcro
®
hook tape.
3/4" x 1-1/4"
Velcro
®
Loop Strip
Back of Monitor
3/4" x 1-1/4"
Velcro
®
Loop Strip
Mounting Clip
Mounting Clip
End View of Monitor