Operating Instructions
Further information (including videos) regarding setting up and running the
multiSUB
®
units can be found at
www.cleaverscientific.com
Usage Guidance and restrictions
Maximum altitude 2,000m.
Temperature range between 4°C and 65°C.
Maximum relative humidity 80% for temperatures up to 31
O
C
decreasing linearly to 50% relative humidity at 40
O
C.
Not for outdoor Use.
This apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664.
POLLUTION DEGREE 2, states that: “Normally only non-conductive pollution
occurs.
Occasionally, however, a temporary conductivity caused by condensation
must be expected”.
Setting up the Horizontal Gel Tanks
Fitting Electrodes
Note the position of the lid on the unit. This shows the correct polarity
and the correct orientation of the cables, black is negative and red
positive.
Remove the lid from the unit. Note if the lid is not removed, fitting the
cables may result in un-tightening of the gold plug and damage to the
electrode.
Screw the cables into the tapped holes as fully as possible so that there
is no gap between the lid and the leading edge of the cable fitting.
Refit the lid.
Fitting Loading Guides
These can be fitted to enhance visibility of the wells if desired. They can be
fitted to the white vinyl platform sheet or to the unit itself.
1. Seat the tray in the unit and note the position of the comb grooves. The
samples run black to red but the trays can be used frontward or
backwards so ensure that the comb grooves closest to the black
electrode are marked.