43
6.3
Troubleshooting
PROBLEM
POSSIBLE CAUSE
SOLUTION
Camera not connected
or not responding
Camera power cable not
plugged in
Software driver for the
camera is missing
Make sure all cables are
connected as shown in the
Installation quick guide
Install Camera driver
Press the refresh button on the
gelLITE icon
‘Hardware not available’
You do not have the
correct filter or lighting for
the
dye
you
have
selected
Check the ‘hardware list’
on the hardware screen
Contact
[email protected]
to upgrade your system
Programme in any extra lights
or filters that you may have on
the hardware screen
Dye not in list
Your dye is not in the
database
Contact
[email protected]
Dye name is greyed out
You do not have the
appropriate hardware to
image that particular dye
Check the ‘hardware list’
on the hardware screen
Contact
‘[email protected]’
to upgrade your system
Programme in any extra lights
or filters that you may have on
the hardware screen
Transilluminator will not
turn on
Make sure the
transilluminator switch is
in the ‘on’ position
Make sure the darkroom
cabinet is completely
closed
If the transilluminator will not
turn on then please contact
[email protected]
Transilluminator will not turn on
when the cabinet door is open
for safety reasons
The tubes may need replacing