How temperature affects %RS?
Saturation begins at the coolest spot, so the temperature differences within the
decontaminated space need to be taken into account. When the event of condensation must
be avoided the sensor is located to the coldest spot where the %RS value would reach a
100 first (see picture 4).
Sophisticated control software of Cleamix VCS-100Cr enables hydrogen peroxide- and water
vapour levels to be maintained below saturation point in various bio-decontamination
situations and risk of hydrogen peroxide condensation accumulating on surfaces can be
minimized.