User Manual  

VCS-100Cr  

Decontamination Unit 

 

 

Compatible for software version 2021-01-02 vs. 2.3 

Содержание VCS-100Cr

Страница 1: ...User Manual VCS 100Cr Decontamination Unit Compatible for software version 2021 01 02 vs 2 3 ...

Страница 2: ...ination unit with hose connector 13 Catalytic Converter 16 HPP272 Sensor 17 Operating with Extension Units 18 Get started 18 Settings 19 Login 19 Self check 19 Supply 20 Set 20 Circulate 20 Cat 21 Dash 21 Extensions and Info 22 Prg 23 Save and Load 24 Delete 24 Enviro 24 Trends 25 Importing of measured data 26 Alarms and notifications 27 Tightvnc viewer installation and configuration guide for con...

Страница 3: ...r is an effective biocide in a broad temperature range Under certain circumstances it can create a condensate with water vapour unfavourable for the supposedly dry decontamination process High humidity rates and or low temperatures are likely to cause condensation This can be prevented by conducting dry heated air through the space prior to hydrogen peroxide vapour production The lower humidity pe...

Страница 4: ... VCS 100ExtU unit is a slave unit that receives commands from the master unit Cleamix VCS 100Ext uses up to 50 aqueous hydrogen peroxide solution for generating hydrogen peroxide vapour capable of destroying even the most resistant microorganisms like bacterial spores at room temperatures and low concentrations Identification data Model of the device operating voltage serial number and manufacturi...

Страница 5: ...dizer Precautionary Statements Prevention P271 Use only outdoors or in a well ventilated area P260 Do not breathe mist vapours or spray P280 Wear protective gloves protective clothing eye protection face protection P210 Keep away from heat sparks open flames hot surfaces No smoking P220 Keep Store away from clothing flammable materials combustibles Make sure there is no hydrogen peroxide liquid le...

Страница 6: ...P310 Immediately call a POISON CENTER or doctor P370 P378 In case of fire Use water for extinction Other information Keep container in a cool place out of direct sunlight Store only in vented containers Do not store on wooden pallets Do not return unused material to its original container Avoid contamination Contamination could cause decomposition and generation of oxygen which may result in high ...

Страница 7: ... if a large amount has been ingested In case of skin contact may cause burns erythema blisters or even necrosis Indication of immediate medical attention and special treatment needed if necessary Hydrogen peroxide at these concentrations is a strong oxidant Direct contact with the eye is likely to cause corneal damage especially if not washed immediately Careful ophthalmologic evaluation is recomm...

Страница 8: ...n and generation of oxygen gas which could result in high pressures and possible container rupture Empty drums should be triple rinsed with water before discarding Utensils used for handling hydrogen peroxide should only be made of glass stainless steel aluminium or plastic Pipes and equipment should be passivated before first use Hydrogen peroxide should be stored only in vented containers and tr...

Страница 9: ...14126 class 6 Pouring liquid hydrogen peroxide from the 20 litre canister into the 1 litre bottle 1 Safety Make sure that you are wearing the proper safety wear non flammable clothing rubber boots protective gloves and protective goggles Make sure that you have plenty of water available for rinsing in case the HP solution gets spilled It s a good idea to keep an emergency kit and eyewash pack near...

Страница 10: ...aisala HPP 272 hydrogen peroxide sensor HYDROGEN PEROXIDE BOTTLE Hydrogen peroxide bottle which is placed to the device ETHERNET CABLE For connecting the catalytic converter master unit and extension units POWER CORD Power cord for the VCS devices TABLET Tablet for remote controlling and monitoring CATALYTIC CONVERTER For vapor removal POWER CORD FOR CAT C Power cord for the Catalytic Converter ...

Страница 11: ...Cr 2021 01 02 vs 2 3 User Manual for VCS 100Cr 10 Technical Specification VCS 100Cr ...

Страница 12: ...Cr 2021 01 02 vs 2 3 User Manual for VCS 100Cr 11 Technical Specification VCS 100Ext ...

Страница 13: ... during the decontamination cycle 10 Keep the area clear from visitors 11 Use warning signs in addition to upfront informing 12 Position the device on a levelled surface in a central location so that the vapour can spread freely into the space 13 Remove cushion from the case it absorbs hydrogen peroxide Power cord Plug in the power cord by pressing and rotating the socket clockwise to lock to the ...

Страница 14: ...er installation is especially important in close loop systems and should not be neglected Always have a H2O2 detector close by Installing the hose connector Take four 4 of the screws part A and corresponding gasket A and install them according to the image 2 in the in and outlets of the gas Make sure to align the gasket A and part A so that no gaps appear and check that the installation is tight a...

Страница 15: ... reference Installing hoses to the hose connector Install the supplied hose to the connector with the bridge clamps supplied in the package Image 1 to part B Install the hose as shown in image 5 Confirm that the hoses are tightly secured by gently pulling them If you have not ordered hoses from Cleamix Oy and are using your own Cleamix Oy is not responsible if gas leakage occurs due to non conform...

Страница 16: ... that it is located in upright position and not in angle Ideal location is on table or other flat object so that the unit is not disturbed or moved by accident If you are not using Cleamix Ltd supplied hoses make sure that the used hoses are made from non porous materials are designed for gas usage and the material is able to withstand 50 hydrogen peroxide vapor Please refer to material compatibil...

Страница 17: ...ter is an air circulating fan which is included with HEPA filter and potassium permanganate filter that makes a chemical reaction with hydrogen peroxide vapour where the vapour decomposes into water and oxygen immediately Connect the ethernet cable between Catalytic Converter and VCS 100Cr or VCS 100ExtU Switch the power on at Catalytic Converter Catalytic Converter ...

Страница 18: ...user serviceable parts inside the sensor body Cleaning and Maintenance You can clean the H2O2 and humidity sensor body and temperature sensor body and cable by wiping them with a soft lint free cloth moistened with water or mild cleaning agent such as isopropanol Do not wipe the filter wiping the filter may block its pores and or deposit residue on the filter If the filter is heavily contaminated ...

Страница 19: ...vice Extension units should be connected like star wired principle i e the cords must not generate a loop Master units must not be connected together Operating with Extension Units Turn the power on the device using the power switch VCS 100Cr devices are equipped with Samsung Galaxy Active tablets for remote control Take the tablet and connect it with the VCS 100Cr device by entering to the tablet...

Страница 20: ...ords Service ID for authorized service This ID level can modify Admin and User level ID s Settings Login When turning on the device there will appear LOGIN PAGE Enter your user name and password If you forget your password you can enter using Default User login by pressing GO Change password by selecting the user and adding the current password Then press Change Password Admin level user can Manag...

Страница 21: ...ify RS value to field RS Limit Set the lowest PPM level where the treatment time will start using the field Start treatment time at The Wi Fi is ON when the background of the button is dark When Wi Fi is disabled the WiFi card is un powered so that the device cannot be hacked This might be needed in high security areas Set Circulate Before start Check the proper connection of the hoses from the HP...

Страница 22: ...tered the device will always start the CAT from the highest value The CAT can also be controlled manually by switching CAT Manual Override ON OFF Dash On DASH page can be monitored Hydrogen peroxide level PPM Relative saturation of water and hydrogen peroxide RS Relative humidity RH Temperature TEMP Remaining content of H2O2 LIQUID Elapsed treatment time TIME Total time TOTAL TIME Possible alarm i...

Страница 23: ...ess by pressing again Pause Extensions and Info page shows the information about the Master unit and the additional Extension Units and CAT s Each Ext has it s own ID The ID s last four numbers are the same as the last four numbers in the Ext serial number that can be found on the device s register plate From the ID you can recognize the external units from each other ID s for the CAT s and the Ex...

Страница 24: ...ck program for further use This function can be used for example in situation where we have area that needs to be decontaminated on regular basis With quick programs we can make the regular basis decontamination sessions faster and more efficient since we don t need to set up the settings every time Prg ...

Страница 25: ...with the set treatment values Press Save and Enter the Recipe Name Load saved recipes Select the saved recipe from list and press Load Delete Delete recipe by selecting the saved recipe and press Delete Enviro Environment page shows the parameters of the surrounding environment ...

Страница 26: ...RY Selects the History mode i e earlier captured data can be viewed and scrolled by left right arrow buttons LEGEND Shows the curve colours and to select which curves are shown Upper graph PPM RS RH Lower graph PPM RH and RS change rate and TEMP RH RS The PPM at upper graph has auto scale the ymax value is increased automatically by steps of 50 If ymin ymax values were changed for manual scale the...

Страница 27: ...an be imported as Excel table or as PDF document Pressing Update time the time on the report filename will be updated When generating the reports must wait that the message on the left upper corner has disappeared If new report is chosen too early the created file might be faulty DEL LOG Press to empty the log DEL RPTS Select the reports to delete from list Then press Delete selected GEN RPTS The ...

Страница 28: ...g working unwell with the unit Alarms can be found on the Alarms page The unit will send notifications if there is something critical that has been overlooked Possible notifications Sensor not connected Supply value is zero Treatment time is zero Target PPM is zero Liquid circulation was not done If the temperature is less than 17 C Warning Temperature is low start treatment anyway The unit will w...

Страница 29: ...tions 256 colors Local Cursor Shape Arrow Press OK and then press Connect This is for Windows only Load the TightVNC installation package from the link below and start the installation https www tightvnc com download 2 8 27 tightvnc 2 8 27 gpl setup 64bit msi Steps Next Accept license Next Choose Custom setup type Do not install Server select Entire feature will be unavailable ...

Страница 30: ...catalytic converter The ventilation of the space must be planned case specifically and in accordance with valid regulations Note that porous materials within the decontaminated space may increase the hydrogen peroxide aeration time Switch off the device detach the power cord by lifting the latch and turning the socket counter clockwise until it gets loose Detach the hydrogen peroxide sensor by lif...

Страница 31: ...k fuses F2 F4 Note If the same fuse blows again the device should be delivered to a country specific authorized distributor or the manufacturer for service Heat plate stops heating In case that the heat plate stops heating the unit must be delivered to a country specific authorized distributor or to the manufacturer The device contains an overheating protection that will be discharged if the surro...

Страница 32: ...an be successfully bio decontaminated by hydrogen peroxide vapour Appendix II Controlling a Cleamix VCS 100Cr bio decontamination process When optimizing a decontamination process the importance of continuous measurement of the bio decontamination process parameters can t be over emphasized The maximum achievable H2O2 vapour concentration is highly dependent on the existing environmental condition...

Страница 33: ...en the event of condensation must be avoided the sensor is located to the coldest spot where the RS value would reach a 100 first see picture 4 Sophisticated control software of Cleamix VCS 100Cr enables hydrogen peroxide and water vapour levels to be maintained below saturation point in various bio decontamination situations and risk of hydrogen peroxide condensation accumulating on surfaces can ...

Страница 34: ...rials challenge the optimal distribution and delivery of the bio decontamination agent Typical non condensing decontamination cycle is divided into three phases Once the targeted parameters of a decontamination process have been attained for a sufficient exposure period the space has to be properly aerated It is possible to enter the space safely only when the H2O2 vapour concentration gets below ...