OE 0900
HGV INDUCTION HEATER
•
"HC" Heating mode
Select the desired heating mode - see chapter Heating modes. The following options are
available:
- CF = control based on the desired magnetic field strength
- CP = control of the desired heating capacity
•
"Po L" Maximum output power
Allows you to select the maximum output power of the inverter, which is suitable for small
coils, limited by the maximum power or current. This allows the use of specific coils, small
or wire, that would otherwise overheat.
The other advantage is the possibility of limiting the maximum power to the minimum
necessary, which, in combination with the CP power control mode, considerably reduces
the overheating of the unit and increases the thermal efficiency.
•
"ic.L" Line current limit
Allows to limit the current consumed by the power line, which allows a trouble-free use of
the device in networks where electrical safety equipment is of poor quality, for example in
repairs, the construction industry, workshops, etc.
Indicator lamp
Statuts
Red
Parameter selection
Green
Value settings
•
The indicator lamp will have the following states:
•
The following parameters can be set:
Description
Display
Value
Return/exit
---
---
Heating mode
REG
CF = constant magnetic field/current
CP = constant heating power
Input Current Limit
ic.L
6.0 max. - 16,5A = limitation on max. input current
Max. output power
Po.L
2.0 max. - 4,5 kVA ; default value : 4 kVA
Cooling
CoL
Lo = silent mode (workshops, short-term capacity)
Std = standard mode
Hi = max. capacity / does not switch off
Water fill
FIL
no
yes = activates filling and venting
Factory setting
FAC
no
yes = load factory setting
Unlock code
UnL
Servicing menu code, to unlock features
Firmware
Fir
Display firmware version