Colibrick A/D Converter
3 Installation
3.2 Standard cable for Clarity station
Fig 4: Cables for the
Clarity
station package
The standard
Clarity
station package includes a cable set composed of
signal, starting and digital output cables for connecting the
Clarity
to the
chromatograph and an
USB
cable for connecting the
Colibrick
to the
computer.
Signal cables
The cables, labeled “
DET1
” to “
DET4
” (according to the number of
channels), are supplied as standard without connectors only with stripped,
tinned endings - red/brown (+), white (-) and shielding (analog ground).
Starting (marker) cables
The cables, labeled “
IN1
” to “
IN4
”(according to the number of channels),
are ended with a RCA connector. Each starting cable has one cable with
free leads [red (+), shielding (digital ground)] for direct connection to the
chromatograph or to a valve and one cable ended with a button for those
cases where there is not a starting contact available and it is necessary to
perform a manual start.
Digital output cables
Relay contacts, labeled "
OUT 1R
" to "
OUT 4R
" (according to the number
of channels), are ended with free leads. They are used for synchronizing
autosamplers in the active sequence without an AS Control module.
At the
Colibrick
end, the cable ends on a female CANNON SUB D 37-pin
connector.