GUS Disinfection Soak Station™
M0036_F
6
G32-S / G32-E
Operation
Placing Instruments into the System
1.
All scopes and instruments should be pre-cleaned and dry prior to soaking in high-
level disinfectant.
2.
Open the front acrylic doors of the system and remove the lid on the soak tray(s).
3.
Gently place your instruments into the tray(s) containing high-level disinfectant.
4.
Follow instrument manufacturer’s instructions for flushing lumens and channels of
scopes.
5.
Replace lid(s) and close acrylic door.
Removing and Rinsing Instruments
1.
When done soaking, open the front acrylic doors and remove the lid from the soak
tray(s).
2.
Gently lift instruments from the soak tray(s) and shake off excess disinfectant.
3.
Place instruments into the rinse tray(s)
inside the GUS
system
to remove residual
chemical.
NOTE:
If three trays are have provided with your system, the rinse water should be
placed in the middle of your GUS station.
4.
Remove from rinse tray(s) and conduct final rinsing and wiping procedures as you
normally do.
5.
Refer to your instrument’s instruction manual for proper rinsing and storage
procedures.
6.
You must leave the system running at all times when the soak tray(s) contains high-
level disinfectant.
Disposal of Disinfectant
In some states and municipalities, it is illegal to dispose of used disinfectants without first
deactivating them. Deactivation also protects the user from exposure to splashes and spills.
CIVCO Medical Solutions manufactures
Glute Out
®
, a glycine-based neutralizer
(part# 610-1045).
After the disinfectant is deactivated, it is safe to dispose of down the sink/drain with cold
water running.
Thoroughly clean and rinse out the container. With the unit turned on, put the soaking
container back into place and refill it with fresh high-level disinfectant.