GUS Disinfection Soak Station™

  

 

M0036_F 

G32-S / G32-E 

 

Operation 

 

Placing Instruments into the System 

 

1.

 

All scopes and instruments should be pre-cleaned and dry prior to soaking in high-

level disinfectant. 

2.

 

Open the front acrylic doors of the system and remove the lid on the soak tray(s). 

3.

 

Gently place your instruments into the tray(s) containing high-level disinfectant. 

4.

 

Follow instrument manufacturer’s instructions for flushing lumens and channels of 

scopes. 

5.

 

Replace lid(s) and close acrylic door. 

 

Removing and Rinsing Instruments 

 

1.

 

When done soaking, open the front acrylic doors and remove the lid from the soak 

tray(s). 

2.

 

Gently lift instruments from the soak tray(s) and shake off excess disinfectant. 

3.

 

Place instruments into the rinse tray(s) 

inside the GUS

 

system

 to remove residual 

chemical.  

 

NOTE: 

 

If three trays are have provided with your system, the rinse water should be 

placed in the middle of your GUS station. 

 

4.

 

Remove from rinse tray(s) and conduct final rinsing and wiping procedures as you 

normally do. 

5.

 

Refer to your instrument’s instruction manual for proper rinsing and storage 

procedures. 

6.

 

You must leave the system running at all times when the soak tray(s) contains high-

level disinfectant. 

 

Disposal of Disinfectant 

 

In some states and municipalities, it is illegal to dispose of used disinfectants without first 

deactivating them. Deactivation also protects the user from exposure to splashes and spills. 

CIVCO Medical Solutions manufactures 

Glute Out

®

, a glycine-based neutralizer  

(part# 610-1045). 

 

After the disinfectant is deactivated, it is safe to dispose of down the sink/drain with cold 

water running.  

 

Thoroughly clean and rinse out the container.  With the unit turned on, put the soaking 

container back into place and refill it with fresh high-level disinfectant.

 

Содержание GUS Disinfection Soak Station

Страница 1: ...olutions All other trademarks are property of their respective owners Printed in the USA Model G32 S Model G32 E Disinfection Soak Stations Operator s Manual CIVCO Medical Solutions 102 First Street South Kalona IA 52247 USA Tel 1 800 445 6741 Fax 1 877 329 2482 Website WWW CIVCO COM ...

Страница 2: ...ation 4 Filling the High Level Disinfection Tray s 5 Operation 6 Placing Instruments into the System 6 Removing and Rinsing Instruments 6 Disposal of Disinfectant 6 Built in Disposal System 7 Filter Life 8 Filter Replacement 8 Cleaning the System 9 Rinse Water 9 Limited Warranty 10 Registration Card 10 APPENDIX A TROUBLESHOOTING 11 APPENDIX B SITE REQUIREMENTS 12 APPENDIX C GUS PRODUCT SPECIFICATI...

Страница 3: ... of the high level disinfectant that you are using for additional safety precautions Warning Symbols The meaning of the labels and symbols that appear on the packaging and or the system are as follows Dangerous electrical voltage Hazardous area Protective Grounding Signal Words The following signal words are used throughout this manual WARNING Indicates a potentially hazardous situation which if n...

Страница 4: ...re1 or as Space Saver models which are designed to fit under an upper wall cabinet The GUSSystem with the standard filter is designed to remove fumes from Glutaraldehyde Ortho phthalaldehyde OPA Hydrogen Peroxide Optional filters are available for Peracetic Acid Formaldehyde Hydrogen Peroxide WARNING Failure to properly follow the instructions in this manual may result in damage to the system and ...

Страница 5: ... the base There must be a clearance of 1 above the unit 3 Plug the system into a standard wall outlet 4 Remove the protective covering from the acrylic doors 5 Open the acrylic doors and place the soak and rinse tray s s inside Mobile models Assembly of Cart Required 1 Remove the cart from the top of the box and place onto a level surface Assemble as directed 2 Gently lift the system out of the pa...

Страница 6: ...for the high level disinfectant the third tray is for clean rinse water The rinse water tray should be placed in the middle of the Gus system WARNING Not all tray s are autoclavable Check before autoclaving 4 Remove the lid off the high level disinfectant tray s and set aside 5 Prior to pouring the high level disinfectant read carefully and follow all label instructions 6 Activation of the high le...

Страница 7: ...chemical NOTE If three trays are have provided with your system the rinse water should be placed in the middle of your GUS station 4 Remove from rinse tray s and conduct final rinsing and wiping procedures as you normally do 5 Refer to your instrument s instruction manual for proper rinsing and storage procedures 6 You must leave the system running at all times when the soak tray s contains high l...

Страница 8: ...plastic suction hose shorter hose with in line filter into the quick disconnect fitting located next to the disposal On Off switch on the front inside of the system Place the other end of the hose into the soak tray s 4 Insert the discharge hose into the quick disconnect fitting located on the left side of the system Place the other end of the hose into the sink or hopper drain and turn on the col...

Страница 9: ...that the filter change take place at the time of replacing the disinfectant 1 Turn the main power switch OFF and disconnect the power cord from the electrical outlet 2 Remove the top louver plate see figure 2 located at the top of the machine by unscrewing the acorn nut with a 5 16 nut driver Note the G32 has a single top louver plate 3 Remove the old filter Put the used filter into a plastic bag ...

Страница 10: ... hard or abrasive materials Doing so will scratch the surface 1 Turn the power switch OFF and disconnect the power cord from the wall outlet 2 To clean metal surfaces Use a clean lint free cloth moistened with a mild detergent solution and water A mild solvent may be used on these surfaces for difficult to clean areas 3 To clean the HDPE plastic side panels use mild detergent solution such as dish...

Страница 11: ...NS OR OTHER LIABILITIES EXPRESSED OR IMPLIED AND THERE ARE NO WARRANTIES WHICH EXTEND BEYOND THE DESCRIPTION ON THE FACE HEREOF WARRANTIES OF MERCHANTABILITY AND OF FITNESS FOR A PARTICULAR PURPOSE ARISING FROM A COURSE OF DEALING OR USAGE ARE SPECIFICALLY EXCLUDED This warranty does not cover conditions and damages resulting from any of the following Improper installation delivery within the faci...

Страница 12: ... result in exposure to vapor when transferring the instrument to the sink for rinsing NOTE If three trays are have provided with your system the rinse water should be placed in the middle of your GUS station 2 The system is being left on at all times while disinfectant is in the soak tray s 3 Lids are placed on the soak tray s when not in use and during the soaking cycle 4 Fresh disinfectant is be...

Страница 13: ...de front panel of the system see figure 3 The switch is illuminated when the system is ON The power cord plugs into a regular wall outlet The fuse is a 250V 3 0 A 5X20 FSF panel mounted The ratings plate and serial number are located on the right side of the system Figure 3 Front Panel of the System Actual Size Model Air Treated CFM FPM w door closed low FPM w door open high Voltage Current w door...

Страница 14: ...GUS Disinfection Soak Station M0036_F 13 G32 S G32 E APPENDIX C GUS PRODUCT SPECIFICATIONS UL Approval USA and Canadian Standards Approved ...

Отзывы: