DESCRIPTION OF DISPLAY MARK

Message

Meaning

Corrective action

Measurement before 
device stabilization.

Wait until all the icons stop 
fl ashing.

The device showing a 
rapid ambient temperature 
change.

Allow the thermometer to 
rest for at least 30 minutes 
at a room temperature 
between 10 and 40°C (50 
to 104°F)

The ambient temperature 
is not within 10 to 40°C 
(50 to 104°F).

The system is not 
functioning properly.

Remove the battery, 
wait for 1 minute, 
and reinsert it. If the 
message reappears, 
contact the retailer where 
you purchased the 
thermometer for service.

(1) Ear  temperature  mode: 

Temperature taken is 
higher than + 42.2°C 
(108°F).

(2) Material  surface 

temperature mode: 
Temperature taken is 
higher than + 80°C 
(176°F).

The thermometer might 
have measured the 
temperature of a nearby 
object that it was not 
designed to measure. If 
the malfunction persists, 
please contact the retailer 
where you purchased the 
thermometer.

(1) Ear  temperature  mode: 

Temperature taken is 
lower than  + 34°C  
(93.2°F).

(2) Material  surface 

temperature mode:  
Temperature taken 
is lower than  - 22°C  
(- 7.6°F).

Device cannot be 
powered on to the ready 
stage.

Change the battery.

Содержание CT830

Страница 1: ...INSTRUCTION MANUAL for DIGITAL EAR THERMOMETER REF CT830 Probe Probe Cap ON MEM button LCD SCAN button Probe lens IDENTIFICATION OF PARTS Fig 1 English 0801...

Страница 2: ...on on or in the ear 4 DO NOT use the ear thermometer if the main body is damaged for example the probe lens is broken The continuous use of a damaged unit may cause injury improper results or serious...

Страница 3: ...and do NOT allow the unit to come into contact with chemical solvents 4 Do NOT drop the ear thermometer and do NOT expose the ear thermometer to heavy shock The ear thermometer is not shock protected...

Страница 4: ...erature so you can compare it to the measurement result About Ear Thermometers All objects radiate heat To measure body temperature this ear thermometer uses an infrared sensor incorporated in its pro...

Страница 5: ...n for about 5 sec and the icon starts to ash indicating the setting mode Fig 5 Fig 5 Date Time 24 12 hour display 2 Press the SCAN button for the 12 24 hour setting mode The icon starts to ash indicat...

Страница 6: ...emove the probe and read the temperature on the display Fig 8 Fig 8 Ear temperature mark NOTE The time and calendar settings should be done before measurement readings are taken The thermometer always...

Страница 7: ...the MATERIAL SURFACE TEMPERATURE scan mode with icon Fig 10 Fig 10 Material surface temperature mark 3 1 Aim the probe torward the material and press the SCAN button and the material surface temperatu...

Страница 8: ...eps to warn the user of a fever MEMORY There are a total of 9 memory slots for recording EAR TEMPERATURE measurements Each memory slot also records the date and time of the measurement Memory slots ca...

Страница 9: ...NG on the display except the time and date When the ON MEM button is pressed NOTHING is displayed except the battery mark blinking 12 times followed by OFF for 1 second and then the display returns to...

Страница 10: ...oo long may cause the probe to make a higher ambient temperature reading This could make the body temperature measurement lower than usual After each use clean the probe and probe lens with a cotton s...

Страница 11: ...sert it If the message reappears contact the retailer where you purchased the thermometer for service 1 Ear temperature mode Temperature taken is higher than 42 2 C 108 F 2 Material surface temperatur...

Страница 12: ...Q4 Is it hygienically safe to use the ear thermometer without the probe cover A4 The probe and probe lens should be kept clean by using a cotton swab or soft cloth moistened with 70 alcohol Therefore...

Страница 13: ...ent Operating temperature range 10 40 C 50 104 F Storage temperature range 20 50 C 4 122 F Transportation temperature range must be less than 70 C Unit size 115 L X 37 W X 29 H mm Weight Approx 60g in...

Страница 14: ...Prob Prob IDE Design and Specification are subject to change without notice CITIZEN is a registered trademark of Citizen Holdings Co Japan CITIZEN es una marca registrada de Citizen Holdings Co Jap n...

Отзывы: