![Citizen CHU305 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/citizen/chu305/chu305_instruction-manual_2609784021.webp)
- Eng 21 -
Español
English
Model Number:
CHU305
Measurement system:
Oscillometric method
Display
Digital display type
Measurement Localization
Upper Arm
Cuff:
Soft cuff
Cuff circumference range:
8 3/4” – 16 1/2” (22 cm – 42 cm)
Cuff pressure:
0 - 300 mmHg
Measurement range:
Pressure
Systolic: 60 – 260 mmHg
Diastolic: 40 – 199 mmHg
Pulse
40 to 180 pulse/min
Accuracy:
Pressure
±3 mmHg
Pulse
±5% of reading
LCD displays:
Pressure
3 digits
Pulse
3 digits
Icons
: Measurement icon
: Ready to inflate indicator
: Battery icon
: Irregular Heartbeat (IHB) Indicator
: Memory icon
: Hypertension Classification Indicator
Switch:
2 (START/STOP, MEMORY)
Inflation:
Automatic Inflation by internal pump
Deflation:
Automatic speed deflation system.
Rated voltage:
6V DC
(
: direct current)
Exhaust:
Electromagnetic quick exhaust valve
Power supply:
4 X 15V
SIZE AA batteries (LR6)
Battery duration:
Approx. 800 times (Alkaline)
Automatic Power Off function:
Approx. 1 min. (after activated)
Main Unit Dimensions:
Approx. 5 4 (W) x 2.1 (H) x 3.7 (D) in
(138 (W) x 54 (H) x 95 (D) mm)
Cuff Size:
Approx. 6.4 × 22.8 in (163 ×580 mm)
Weight:
Unit: Approx. 7.3 oz. (207g) w/o batteries
Cuff: Approx. 4.7 oz. (134g)
Operating conditions:
Temperature
50°F – 104°F (10°C – 40°C)
Humidity
90%RH
Storage conditions:
Temperature
-4°F – 131°F (-20°C – 55°C)
Humidity
90%RH
Electric shock protection:
Internal power unit
Degree of Protection
Type BF applied part
Mode of operation:
Continuous operation
Memory:
1 x 99 readings, Average of last 3 readings
Service Life:
10,000 measurements or 3 years
Cuff’s Life:
Approx. 1000 times
Protection against ingress of water:
IPX0
Accessories:
Set includes an cuff, 4 AA batteries (LR6) for the
monitor, Instruction Manual/QSG/LOG/FAQ
* Applied part for this device is Cuff.
* A range in barometric pressure 105 kPa to 80 kPa
* A range in altitude -300 m to 2000 m
* The Electronic Sphygmomanometers corresponds to the below standards: IEC 60601-1:2005/EN
60601-1:2006/AC:2010 (Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance), IEC 60601-1-2: 2007/EN 60601-1-2:2007/AC:2010 (Medical
electrical equipment – Part 1-2: General requirements for basic safety and essential performance
- Collateral standard: Electromagnetic compatibility - Requirements and tests), EN 1060-1: 1995
+ A1:2002 + A2: 2009 (Non-invasive sphygmomanometers - Part 1: General requirements),
EN 1060-3:1997 + A1: 2005 + A2: 2009 (Non-invasive sphygmomanometers - Part 3:
Supplementary requirements for electro-mechanical blood pressure measuring systems). IEC
80601–2–30: 2009+Cor. 2010 (Medical electrical equipment – Part 2–30: Particular requirements
for the basic safety and essential performance of automated non-invasive sphygmomanometers)
SPECIFICATIONS
00̲CHU305̲USA.indb Eng-21
00̲CHU305̲USA.indb Eng-21
2015/01/13 19:53:22
2015/01/13 19:53:22