7.7.
Fixed Trigger Shutter Mode
□
This is the mode to start exposure with external input trigger signals, and set the exposure time with
serial commands.
□
Trigger operation is H Sync. V-Sync Reset. The delay time (Exposure Time Delay) from
detecting trigger edge in the camera to starting exposure is max. 1HD.
□
Triggers can be accepted even when outputting video signals. However, trigger signals for
exposure to start the next video output prior to the completion of video transmission for the prior
video output signals cannot be accepted.
□
If the next trigger is input prior to the completion of video transmission for the prior video output
signals, the image may be impacted.
□
Trigger input during exposure time should be ignored. (Refer to the below A)
.