Cirs ZERDINE Inside 055A Скачать руководство пользователя страница 7

4

GENERAL GUIDELINES FOR PERFORMING MEASUREMENTS

It is recommended that all measurements be performed at the most frequently used 
imaging arrangements. The importance of these tests is to make sure the system 
performance remains constant over an extended period of time. Measurements 
may also be used to compare the performance of various setups of the same ma-
chine or to compare different machines in a quantitative manner.

The following are general steps for imaging all targets:

• 

Some wires will appear as short lines rather than dots. When using the  

 

 

electronic calipers, always take measurements from a point on one echo  

 

 

to the same point on the next, i.e., center to center. Otherwise, errors may    

 

be introduced.

• 

If a convex probe is used, center the target within the scan plane in order to   

 

minimize degradation and distortion introduced on the outer edges of the    

 probe.

• 

When assessing vertical distance measurements, DO NOT press on the  

 

 

scanning surface. Pressure on the scanning surface causes the wires to  

 

become temporarily displaced, making vertical distance measurements inac-  

 curate.

• 

When assessing horizontal distance accuracy, ensure the scan plane is  

 

 

perpendicular to the horizontal target group. Rotation of the probe will result   

 

in inaccurate distances.

• 

Always be sure the phantom is scanned while at room temperature. A  

 

 

phantom just received may be colder or hotter than room temperature de-   

 

pending on where it was stored during shipping. Temperature affects the  

 

 

speed of sound and, ultimately, the perceived measurements. The phantom   

 

should be stored at room temperature for at least 24 hours before use to  

 

 

ensure its core temperature is correct.

• 

The most accurate measurements will be made with the phantom 22˚C ±    

 

1˚C (70˚F–73˚F).

ESTABLISHING A BASELINE

Before performing routine quality assurance measurements, establish:

1.  System settings for each measurement:

  System setup can have a dramatic impact on the results obtained from quality as- 
  surance measurements.  You must establish and record what system settings    
  should be used for each of the quality assurance tests.  These same settings  

 

     should be used each time the test is performed.  If not, then the conclusions  

 

  drawn may not be valid. CIRS recommends that you use the most commonly set-      
  tings for the type of probe tested- i.e. the liver preset values for an abdominal    
  probe- which are called a "normal" technique in the sections that follow.

Содержание ZERDINE Inside 055A

Страница 1: ...IDE U LTR A S O UND Q U ALITY ASSURANC E ZERDINE Inside A registered trademark of CIRS Three Dimensional Wire Test Object Model 055A 900 Asbury Ave Norfolk Virginia 23513 USA Tel 757 855 2765 WWW CIRS...

Страница 2: ...2...

Страница 3: ...OF REMOVABLE WATER WELL AND COVERS 3 GENERAL GUIDELINES FOR PERFORMING MEASUREMENTS 4 ESTABLISHING A BASELINE 4 3 TESTING PROCEDURES 5 DISTANCE 5 PERIMETER AREA AND VOLUME MEASUREMENTS 6 UNIFORMITY T...

Страница 4: ...f speed attenuation and relative contrast In addition two ultrasound systems are used to visually inspect each phantom As a result every ultrasound phantom is subjected to rigorous testing both during...

Страница 5: ...tion always store the phantom in the air tight carry case with the removable storage cover attached Inspect your phantom regularly for signs of damage and weight loss If any notice able changes to the...

Страница 6: ...filled with water to provide better coupling Side Fire transducers can be particularly chal lenging to scan with a standard phantom CIRS has designed a removable endo cavity cover for these transduce...

Страница 7: ...sing horizontal distance accuracy ensure the scan plane is perpendicular to the horizontal target group Rotation of the probe will result in inaccurate distances Always be sure the phantom is scanned...

Страница 8: ...system is 10 for distances up to 2 cm the scanner may detect 2 0 cm as being any where from 1 8 cm to 2 2 cm and still be functioning properly The user is responsible for establishing action levels 4...

Страница 9: ...se on subsequent testing 4 Align the probe so that all the horizontal targets are displayed at their maximum intensity level 5 Freeze the image and obtain a hard copy 6 Using electronic calipers measu...

Страница 10: ...ality assurance record These settings should be used for subsequent testing 4 Position the transducer over the line target group until a clear image is ob tained 5 Freeze image and obtain a hard copy...

Страница 11: ...s gain TGC output etc as for a normal technique Record these settings for use on subsequent testing 4 Align the probe so that all the vertical targets are displayed at their maximum intensity level 5...

Страница 12: ...6 3 cm 0 cm x 0 25 cm y 0 5 cm 7 3 cm 0 cm x 0 25 cm y 0 5 cm PHANTOM HOUSING Material 1 4 Black ABS Dimensions Outer 18 cm x 13 cm x 11 cm Dimensions Inside 11 5 cm x 16 5 cm x 11 cm Scanning Surfac...

Страница 13: ...without subsequent damage to the material At normal room temperatures Zerdine will accurately simulate the ultrasound characteristics found in human liver tissue Specific proprietary fabrication proce...

Страница 14: ...utor prior to returning the product Visit https www cirsinc com distributors to find your local distributor If you purchased your product direct through CIRS call Customer Service at 800 617 1177 emai...

Страница 15: ...12...

Страница 16: ...erized Imaging Reference Systems Inc has been certified by UL DQS Inc to ISO 13485 2016 Certificate Registration No 10000905 MP2016 COMPUTERIZED IMAGING REFERENCE SYSTEMS INC 900 Asbury Ave Norfolk Vi...

Отзывы: