What are we going to do

here?

In this section, we’ll give you a step-by-step procedure for installing the Alarm Box 
Add-On accessory on your Cirris 1000H

X

.  To make things easier, we’ll provide 

digital photographs taken at the workbench, as one of our factory technicians 
installed an alarm box.

Procedure

To install the Alarm Box Add-On accessory, follow these steps:

1.

Open the quarter-turn fasteners, and remove the scanner cover plate, then 
gently lift the scanner assembly out of the chassis of the .

In this view, the quarter-turn fasteners which hold the scanner assembly

Scanner assembly

Ribbon Cable
(still connected)

Center Cover
Plate (still attached)

on the analyzer have been opened, and the cover plate has been 
removed.  The scanner assembly has been lifted from the chassis. Note
that the ribbon cable which connects the scanner with the rest of the
analyzer is still connected.  The center cover plate is still in place on 
the analyzer’s chassis.

Installation 
Procedure

Содержание 1000HX

Страница 1: ...Technical Note Installing the Alarm Box Add On on the Cirris 1000HX Cable Analyzer Version 3 0 September 1998...

Страница 2: ...Add On on the Cirris 1000HX Cable Analyzer Version 3 0 September 1998 Copyright 1998 by Cirris Systems Corporation All Rights Reserved Cirris Systems Corporation 1991 Parkway Boulevard Salt Lake City...

Страница 3: ...l need To install the Alarm Box Add On accessory on your analyzer you ll need these things Cirris 1000HX Cable Analyzer Complete Cirris Alarm Box Add On accessory This manual How to contact us If you...

Страница 4: ...Introduction Alarm Box Add On Tech Note 2...

Страница 5: ...en the quarter turn fasteners and remove the scanner cover plate then gently lift the scanner assembly out of the chassis of the 10000H In this view the quarter turn fasteners which hold the scanner a...

Страница 6: ...on the side of the Alarm Box will fit into the holes in the side of the analyzer itself 3 Remove the center cover plate from the 1000HX and set the plate aside Ribbon cable now disconnected Scanner A...

Страница 7: ...ove the processor at an angle as shown This will avoid snapping the pogo pin that grounds the processor to the chassis Note the ribbon cable connected to the Processor 5 Turn the Processor Assembly ov...

Страница 8: ...of the analyzer 7 Gently lift the analog assembly up and out of the chassis The chassis will now be essentially empty Ribbon Cable disconnected Speaker Wires disconnected In this view the Processor A...

Страница 9: ...of both boxes if they re there snake the long ribbon cable through the rectangular openings in the sides of both chassis then butt the two chassis together the pins on the side of the Alarm Box shoul...

Страница 10: ...ribbon cable still aren t connected In the lower photo with the analog assembly in place note how the long ribbon cable and the speaker wires run down alongside the analog assembly Analog Assembly Sp...

Страница 11: ...ct the speaker wire as shown here By now all three cables the wide Analog to Processor ribbon cable the narrower long ribbon cable and the speaker wires should all be connected to the processor assemb...

Страница 12: ...o the processor assembly turn the processor assembly right side up and carefully lower it back into the chassis as shown here Be sure to lower the assembly at an angle as shown This will avoid snappin...

Страница 13: ...the Processor to Scanner ribbon cable then lower the scanner assembly carefully back into the chassis as shown here 16 Replace the scanner cover plate The installation of the Alarm Box is now complete...

Страница 14: ...Installation Procedure 1000HX Alarm Box Add On Tech Note page 12...