Active Periodic Cooling (APC) Menu
50
%
T
H
U
D
37.7
oFF
%
T
H
U
D
cIF
coF
%
T
H
U
D
cb
24
%
T
H
U
D
-8.0
%
T
H
U
D
-Press U + D together / 3s to enter the menu
-Use T /H buttons to pass next / before parameter.
-Use U/D buttons to change parameter values.
cIF: Active Periodic cooling(APC) funciton activation.
oFF: Active Periodic Cooling is deactivated.
cLt:
“cooling for a time” mode.
Stops heater and humidifier
for “hct” time. ( check hct parameter)
cLS: “
cooling for a set ”mode.
Stops heaters and humidifier up
to a adjusted temperature set point and keep temperature at
this point for an adjusted time. (check cSt/FSt)
cb: Active Periodic cooling Starting.
con: first cooling starts immediately after save the menu.
coF: first cooling starts after periodic cooling cycle (cPr)
cPr: Periodic cooling cycle time( hours)
6-8-12-24-48-72 hours selectable.
System repeats cooling in this cycle.
cSt: cooling set point ( for C display models)
cooling set value: temperature set point – cst ( - 8 C)
note: cSt is only active in clS mode on C display models
Example: for a system working on 37.7 C, system
cools up to 29,7 C (37,7 – 8.0)
FSt: cooling set point F. ( for F display models)
cooling set value: temperature set point – FSt (-15 C)
note: FSt is only active in clS mode on C display models
Example: for a system working on 99.8 F, system
cools up to 84,8 F (99,8 – 15.0)
cth: waiting time in cooling set point.( minute)
Waiting time of system in cooling set point
cth is only active in clS mode
cPr
cSt
-15.0
%
T
H
U
D
FSt
30
%
T
H
U
D
cth
180
%
T
H
U
D
hct
on
%
T
H
U
D
chu
10
%
T
H
U
D
80
%
T
H
U
D
hct: max cooling time ( minute)
Heater /humidifier cut off time for clt mode.
Maximum cooling time for clS mode.
chu: humidification activation during cooling
( only for models equipped with humidifier)
on: humidification activated
oFF: humidification deactivated
cht :Humidification time in cooling ( minute)
( only for models equipped with humidifier)
Humidification will start before cooling finished for cht
time.
Example. If cooling will finish within 30 minutes.
Humidification starts after 20 minute(30-10). (see page 21)
chh : max humidification set in cooling( RH%)
( only for models equipped with humidifier)
Humidifier works up to this set and keeps incubator in this
point during humidification period.
cFn : cooling fan activation
( only for models equipped with cooling fan)
On: cooling fan is activated during cooling.
OFF: cooling fan is deactivated during cooling.
-Press T+ H together / 3s to save the changing’s.
During cooling, “cool” blinks on up screen,
CAUTION!
don’t cut the power of incubator
during cooling. Power cut will terminate cooling but
the time for cooling period will continue.
cht
chh
on
%
T
H
U
D
cFn
donE
%
T
H
U
D
SAVE
42
%
T
H
U
D
>Cool<
PD SERIES