BATTERY INSTALLATION
1). Open the battery cover and install four long
life ‘AA’ type batteries into the compartment
as indicated. Make sure that the batteries’
polarities (+/-) are positioned correctly.
Never use force to remove the battery cover.
2). Close the battery compartment cover.
• Replace the batteries when battery replacement indication appears
on the display or if the display is blank when the O/I button is pressed.
• The batteries delivered in the device are intended to check the device
after sale; the period of service of these batteries can on occasion be
less than other high power batteries.
• Replace all batteries simultaneously. Do not recharge batteries.
• If the device is not used for a long period, remove the batteries.
• Do not leave run down batteries in the device.
CORRECT POSITION FOR MEASUREMENT
1). Sit at a table and let it support your arm as you take
the arterial blood pressure measurement. Make sure
that the measurement location on the upper arm is
at approximately the same height as the heart, and
that the forearm is extended naturally on the table
and does not move.
2). You may lie on your back and take the
measurement. Look at the ceiling, stay calm,
and do not move your neck or body during
the measurement. Make sure the measurement
location on the upper arm is at approximately
the same height as the heart.
5
ERROR DISPLAYS
Indication
Possible Reason
The cuff is not correctly
applied or the tube plug
is not inserted tightly.
The measurement could
not be made because of
moving an arm or hand,
or talking during
measurement.
The cuff was not inflated
to the necessary pressure.
Arrhythmia (irregular
heart beat).
The batteries are weak.
Correction Methods
Make sure that the cuff is put on
correctly and the tube plug is
inserted tightly and repeat the
measurement.
Repeat the measurement by
following the instructions.
Repeat the measurement
inflating the cuff to 30-40 mm
Hg above expected systolic
pressure.
Consult your healthcare
professional.
Replace all 4 batteries with
new ones.