2. Specifications
Display:
4 1/2 digit liquid crystal display (LCD) with a maximum reading of 19999.
Over range:
(OL) or (-OL) is displayed.
Low battery indication:
The “
” displayed and blinking when the battery voltage drops
below the operating level.
Measurement rate:
10 times per second (nominal).
Multiple signal measurement:
6500 samples with real-time data. (Max.)
ALM Record:
5400 sample with start/stop time. (Max.)
Single signal measurement:
32000 records capacity. (Max.)
Operating Environment:
0°C to 40°C at < 70% relative humidity.
Storage Temperature:
-20°C to 60°C, 0 to 80% R.H. with battery removed from meter.
Safety:
According to EN61010-1 protection class II over voltage category (CAT II 600V)
pollution degree 2.
Power:
Single standard 9-volt battery, NEDA 1604, JIS 006P, IEC 6F22.
Battery life:
40 hours typical with carbon-zinc.
Dimensions:
192mm (H) x 91mm (W) x 52.5mm (D)
Weight:
335g including battery.
Accessories:
1. One pair test leads
2. Two K-type thermocouples
3. One spare fuse installed (500mA/250V)
4. One RS232 transmission cable
5. One CD-ROM (software)
6. One Adapter
7. 9V battery (installed) and Operating Instructions
8. Manual
Stated accuracy at 23°C ± 5°C, <75% relative humidity.
DC VOLTAGE
Range: 200mV, 2000mV/ADAPTER, 20V, 200V, 600V
Resolution: 10uV
Accuracy: ±(0.05%rdg + 5dgts)
Input impedance: 10M
Ω
Overload protection: 600VDC or AC rms
AC VOLTAGE (50Hz - 500Hz)
Range: 200mV, 2000mV/ADAPTER, 20V, 200V, 600V
Resolution: 10uV
Accuracy:±(0.75%rdg + 4dgts) on 200mV to 200V ranges
±(2.0%rdg + 4dgts) on 600V range
Input impedance: 10M
Ω
Overload protection: 600VDC or AC rms
15
6.2.3 A-H (Auto Hold)
In “Shift Mode”, press “
H
HOLD” button to change the operation mode into “Auto Hold”
Mode.
The occasions of Auto Hold:
1. Auto Hold will not work and “H” blink when the measurement value lower then 50 counts
or OL.
2. When the new Auto Hold value has produced and the value higher the last one than 50
counts then the new one will be updated vice versa.
3. Auto Hold value will be updated when the variation of the value under 10 counts for five
times.
4. Beeper sounds two times and Hold signal displayed when new Auto Hold value produced.
6.2.4 LOG
Stand-By for Data Logging:
In “Shift Mode”, press this button to enter “Data Logging Function”. An indicator “GRP
MEM” and a number show at lower-left corner of LCD, telling how many groups of data had
been logged in the memory.
And the indicator “LOG” shows to the right of the number telling the meter is now in the
“Data Logging Function”.
Note:
1. When the “
” is blanking, the Data Logging function is disabled. Please replace the
battery to remain the Data Logging function (ref. 8).
2. It must connect battery when you use DC transformer as main power to run datalog
function.
Start and stop Data Logging:
Once entering the “Data Logging Function”, press this button to start logging data.
Indicator “LOG” blinks to tell the meter is logging the measured data with a preset time
interval.
And indicator “GRP MEM” is substituted by SAVE Log and the number of data group stored
in the memory.
6.2.5 SAVE
In “Shift Mode”, press this button to enter “SAVE function”. An indicator “GRP MEM” and a
number show at lower-left corner of LCD, telling how many data had been saved in the
memory. And the indicator “SAVE” shows to the right of the number telling the meter is now
in the “SAVE function”.
Note:
1. When the “
” is blanking, the Data Logging function is disabled. Please replace the
battery to remain the Data Logging function (ref. 8).
2. It must connect battery when you use DC transformer as main power to run datalog
function.