Operation Overview
3-5
the load terminal to perform the measurement.
4
V/I Mon: Two separate BNC connector to simulate the load voltage and
current. VMON is 0~10V that map to 0V~full scale voltage while IMON is
0~10V that map to 0A~full scale current.
5
System Bus: The connectors for connecting multiple 63200A Series Load
in parallel or series.
6
DIGITIAL I/O: The connector for external waveform input and digital
system input/output signals. The digital system input/output signals are TTL
compatible.
7
Extended communication interface: GPIB interface for extension.
8
USB Device: It connects the PC and remote controller.
9
The power fuse.
Voltage & Current Monitoring (V/I Mon)
3.3.1
Each channel on the load has two independent BNC connectors to monitor the voltage and
current, also to output signals to I MON and V MON. The connectors are located on the
rear panel. A 0 to 10V output signal is mapping to a 0 to full scale input range.
System Bus Port
3.3.2
A System Bus is a common used parallel port for 63200A Series Electronic Loads. The two
System Bus ports are 10-pin connectors (RJ-45 male connector.)
Be sure to use the cable
of Chroma’s standard accessory and ensure the load input power is connected correctly
before connecting the System Bus.
See section 3.5.3 for the detailed parameter settings of System Bus.
Содержание 63200A Series
Страница 1: ...High Power Electronic Load 63200A Series Operation Programming Manual ...
Страница 2: ...High Power Electronic Load 63200A Series Operation Programming Manual Version 1 0 March 2015 ...
Страница 7: ...vi ...
Страница 11: ......
Страница 15: ......
Страница 22: ...Overview 1 7 Figure 1 3 Dimension of 63200A with Protection Cover ...
Страница 23: ......
Страница 31: ......
Страница 127: ......