Operation Overview
3-11
AUTO ON, set for auto loading at power on
When Auto is on, the Load will apply the loading parameters and mode set last time before
turned off for loading when power on next time. The default is OFF.
Measurement
3.5.2
Window Time
This function adjusts the average measurement time.
The setting range is 0.001s~10s and the default is 0.02s.
Sign of Voltage
This function changes the voltage sign for display.
The voltage shows a negative sign when MINUS is selected and shows no sign if PLUS is
selected. The default is PLUS.
Digitizing (capturing measured data)
The 63200A Series Electronic Load provides data capturing function for measured data to
record the waveform. It can record the measured data during loading via this function.
Parameters:
Sampling Time: The sampling time for measured data.
Sampling Point: The total sampling point for measured data.
Trig Source: The trigger conditions for data capturing. There are Load ON, Load OFF, TTL
(DIGITAL IO:TRIG_DIGI signal), BUS trigger and Manual trigger available for
triggering. The default is Load ON.
Trig Point: Set the trigger point.
DIGITIZING: Trigger the data capturing.
Parallel & Sync.
3.5.3
The Electronic Load is able to set for parallel and synchronization. First follow the steps
described in section 2.4.3 to connect the SYSTEM BUS on the rear panel.
Содержание 63200A Series
Страница 1: ...High Power Electronic Load 63200A Series Operation Programming Manual ...
Страница 2: ...High Power Electronic Load 63200A Series Operation Programming Manual Version 1 0 March 2015 ...
Страница 7: ...vi ...
Страница 11: ......
Страница 15: ......
Страница 22: ...Overview 1 7 Figure 1 3 Dimension of 63200A with Protection Cover ...
Страница 23: ......
Страница 31: ......
Страница 127: ......