Programmable Bidirectional DC Power Supply 62000D Series
Operating and Programming Manual
3-22
A1 A2 A3 A4 A5 A6 A7 A8
B1 B2 B3 B4 B5 B6 B7 B8
All New vaules
buffer
buffer
Next time
Figure 3-33
2.
Assuming the Reading Average Time is set to = 8, Average Method
to MOVING, the readings sampling is that the device removes the
oldest sample in the buffer and saves a new sample, then average
them in repetition as Figure 3-34 shows.
A1A2 A3A4 A5 A6 A7A8
buffer
buffer
Next time
A1
A2 A3A4 A5 A6 A7 A8 A9 A10
Figure 3-34
3.
The panel reading is refreshed in the rate of 200ms.
Following is the way to change the Average Method:
4.
Tap the touch screen to select the desired average method. Average Method can be set
to FIXED and MOVING as shown in Figure 3-35.
Figure 3-35
3.2.3.6
Screenshot
In Factory Default, tap Screenshot to turn on or off the screen capture gadget. It allows the users
to capture the desired screens such as protection or error messages and save them to USB for
technical service or RD engineer use. The screen shows as Figure 3-36 (with a small camera icon
appears on the lower right corner.)
Содержание 62000D Series
Страница 1: ......
Страница 2: ...Get more information by downloading Chroma ATE Solutions APP ...
Страница 8: ...vi ...
Страница 9: ...vii ...
Страница 10: ...viii ...
Страница 14: ......
Страница 18: ......
Страница 35: ...Overview 1 17 Figure 1 6 Dimension of 62000D 600 1200 1800 Models ...
Страница 36: ......
Страница 52: ......
Страница 54: ...Programmable Bidirectional DC Power Supply 62000D Series Operating and Programming Manual 3 2 Figure 3 2 ...
Страница 63: ...Manual Operation 3 11 Figure 3 13 Figure 3 14 ...
Страница 109: ...Manual Operation 3 57 _EXT ON SIGNAL EXT ON DISABLE EXT ON ENABLE Output Output Figure 3 85 ...
Страница 110: ......
Страница 117: ...Program Sequence 4 7 Figure 4 8 Figure 4 9 ...