Manual Operation
3-55
3.2.4.5
External Output
The setting of DI1 and DI2 function needs to work with external Analog Interface as Figure
3-82 shows.
Figure 3-82
1.
Remote INHIBIT function allows users to remotely turn-off power supply. Logic: Low
sets Remote INHIBIT to enable
, the power supply’s ON/OFF is still controlled by “
”.
When the Pin 6 and Pin18 (_INHIBIT) of ANALOG INTERFACE are triggered in low
level
that equals to tap “
” on the front panel and set OUTPUT = OFF. The DC
power supply will shut down and sends out a protection signal (in this case the
“
”
button will be off), and Pin6 and Pin18 (_INHIBIT) of ANALOG INTERFACE cannot be
used to release the protection.
2.
When protection occurs to REMOTE INHIBIT the main page will appear the protection
message as Figure 3-83 shows.
3.
Tap
“
Confirm
” to return to the main page.
Figure 3-83