HIPOT Tester 19071/19072/19073 User’s Manual
5-20
DC mode
Item Number
Name
Size (byte) Unit
Description
1
Mode
1
-
2:DC Mode
2
Meter 1 (Source)
2
V
30000: Maximum, 31000: Not Value
4
Meter 2 (Current)
4
100nA 1000000000: Maximum, 1100000000: Not Value
8
Meter 3 (Inrush)
4
100nA 1000000000: Maximum, 1100000000: Not Value
16
Ramp Time
2
100mS
30000: Maximum, 31000: Not Value
32
Dwell Time
2
100mS
30000: Maximum, 31000: Not Value
64
Test Time
2
100mS
30000: Maximum, 31000: Not Value
128
Fall Time
2
100mS
30000: Maximum, 31000: Not Value
IR mode
Item Number
Name
Size (byte)
Unit
Description
1
Mode
1
-
3:IR Mode
2
Meter 1 (Source)
2
V
30000: Maximum, 31000: Not Value
4
Meter 2 (Resistance)
4
100kOhm 1000000000: Maximum, 1100000000: Not Value
8
Meter 3
4
-
Reserved
16
Ramp Time
2
100mS
30000: Maximum, 31000: Not Value
32
Dwell Time
2
100mS
30000: Maximum, 31000: Not Value
64
Test Time
2
100mS
30000: Maximum, 31000: Not Value
128
Fall Time
2
100mS
30000: Maximum, 31000: Not Value
GC mode
Item Number
Name
Size (byte)
Unit
Description
1
Mode
1
-
4:GC Mode
2
Meter 1 (Source)
2
mA
30000: Maximum, 31000: Not Value
4
Meter 2 (Resistance)
4
100mOhm 1000000000: Maximum, 1100000000: Not Value
8
Meter 3
4
-
Reserved
16
Reserved
2
-
Reserved
32
Dwell Time
2
100mS
30000: Maximum, 31000: Not Value
64
Reserved
2
-
Reserved
128
Reserved
2
-
Reserved
PA mode
Item Number
Name
Size (byte)
Unit
Description
1
Mode
1
-
5:PA Mode
2
Under Test Signal
2
-
1:Off, 2:On
4, 8, 16, 32,
64 or 128
Message
16
-
C String, maximum length is 15
OS mode
Item Number
Name
Size (byte)
Unit
Description
1
Mode
1
-
6:OS Mode
2
Meter 1 (Source)
2
V
30000: Maximum, 31000: Not Value
4
Meter 2
(capacitance)
4
pF
1000000000: Maximum, 1100000000: Not
Value
8
Reserved
4
-
Reserved
16
Reserved
2
-
Reserved
32
Reserved
2
-
Reserved
64
Test Time
2
100mS
30000: Maximum, 31000: Not Value
128
Reserved
2
-
Reserved
Return data: The flag of new test result (1 byte) + serial number of step (1 byte) + test
result (1 byte) + measurement value option (1 byte) [ + measurement
value 1 + measurement value 2 + …]