Introduction
1-3
1.3
Specifications (18ºC
∼
28ºC RH
≤
70%)
Withstanding Voltage Test
Test Voltage
AC: 0.05
∼
5kV / DC: 0.05
∼
6kV Constant Voltage (19072,
19073 only)
Voltage Regulation
≤
1%+5V, Rated Load
V-display Accuracy
± (1% of r 5 counts), 2V resolution
Cutoff Current (Note1)
AC: 0.1mA ~ 20mA, DC: 0.01mA ~ 5mA (Note1), 0.1uAdc
resolution
I-display Accuracy
0.1uA-299.9uA
0.3mA-2.999mA
3mA-20mA(5mAdc)
± (1.5% of r 5 counts) (Note2) WDC only
± (1.5% of r 5 counts) (Note2)
Output Frequency
50Hz, 60Hz
Test Time
0.1 ~ 999 Sec, continue (Note3)
Dwell Time
0.1 ~ 999 Sec, off
Ramp Time
0.1 ~ 999 Sec, off (0.1sec approx.)
Fall Time
0.1 ~ 999 Sec, off
Arc Detection
Setting Mode
Programmable Setting
Detection Current
AC: 1mA ~ 20mA, DC: 1mA ~ 5mA
Min. pulse width
10us approx.
GOOD/NO-GO Judgment Function
Judgment System
•
Window comparator
•
A NO-GO judgment is made when a current greater than
the high limit value or smaller than the low limit value is
detected.
•
When a NO-GO judgment is made, the output voltage is
cut out and a NO-GO alarm signal is delivered.
•
If no abnormal state is detected during the test time a
GOOD Judgment is made and a GOOD signal is
delivered.
Insulation Resistance Test (19073 only)
Test Voltage
DC: 0.05kV ~ 1kV, Constant Voltage
V-display Accuracy
± (5% of r 5 counts) (open voltage), 2V resolution
Resistance Range
0.1 M
Ω
∼
50 G
Ω
(Note 4)
Measuring Accuracy
≥
500V: 1M
Ω
∼
1000M
Ω
: ± 4% + 5 counts
1G
Ω
∼
10G
Ω
: ± 7% + 5 counts
10G
Ω
∼
50G
Ω
: ± 12% + 5 counts
< 500V: 0.1 M
Ω
∼
1000M
Ω
: ± 7% + 5 counts
< 100V: 0.1 M
Ω
∼
1000M
Ω
: ± 10% + 5 counts
Secure Protection Function
Fast Output Cut-off
Approx. 0.4mS, after NG happen
Fast Discharge
Approx. 0.2S, Typical
Ground Fault Interrupt
0.5mA ± 0.25mAac (ON), OFF
Continuity Check
0.1
Ω
~ 5.0
Ω
± 0.2
Ω
, GC MODE
Panel Operation Lock
YES
Memory Storage