Electrical Safety Analyzer 19032-P Quick Start Guide
11
voltage. Please connect the cover of the analyzer to earth firstly
when high voltage output terminal HV1 is short-circuited with HV2
terminal.
<<< Dangerous Event >>>
13. The danger management
Under any danger circumstances, such as shock, DUT burning or the
main unit burning. Please obey the following procedures to avoid the
more danger.
•
Cut off power switch firstly.
•
Then pull off the plug of power cord.
<<< Solution >>>
14. Problems
Under the below circumstances, the occurred problem are very danger.
Even press [STOP] key, the output terminal may output high voltage.
•
When press [STOP] key, DANGER indication LED is still light.
•
The voltage meter without voltage reading but DANGER LED is still
light. When the above conditions are occurred, please turn off
power and pull off AC power plug immediately. Don’t use any more,
please send to our company or office for reparation.
15. DANGER indication LED error
When press [START] key, there is already reading on the voltage
meter and DANGER LED is still not light. In the meantime, the
indication LED may be error please turn off immediately. Please send
it to our company or office for reparation.
16. If the analyzer needs long time using under normal operation.
Please notice the following items.
If the high limit setting value is 100.0mA(withstand voltage test), please
notice its ambient temperature. When the ambient temperature is
higher than 40
°
C, please stop operation until it cools down to normal
temperature.
17. The used AC INPUT power of analyzer is
90Vac ~ 250Vac, 50 or 60
Hz.
Only can replace fuse under power-disconnected status, remove fuse
stand from power socket and press new fuse slightly into fuse stand
then plug in the power socket.