Rotational Vacuum Concentrator RVC 2-18 CDplus
HCl resistant version
11 Appendix
Version 04/2012, Rev. 2.3 of 03/07/2017
• sb
73 / 84
Translation of the original operating manual
11.3 Brief operating instructions
1.
Switch the system on via the mains power switch on the right side of the
unit.
2.
Switch the vacuum pump and/or the cooling trap on.
3.
Close the lid.
4.
Starting the warm-up phase:
• Enter the set values for the warm-up phase under "menu"
"options"
"timer warm-up" and "temperature warm-up". Confirm
your entries and quit the menu.
• Press the left-hand function key "mode".
• Select the "warm-up" menu and confirm.
• Open the lid when the warm-up phase has ended.
• Install the rotors and close the lid.
5.
Changing the set values for time, temperature, vacuum and safety
pressure:
• Select "menu" with the right-hand function key.
• Select "change set values for manual mode" from the main menu.
• Select the set value with the aid of the up/down keys and confirm
with the right-hand function key "edit".
• Change the set value with the aid of the up/down keys and confirm
with the right-hand function key "ok".
• Quit the menu with the left-hand function key.
6.
Starting the process "evaporation manual":
• Open the "mode" menu with the left-hand function key.
• Select "evaporation manual" using the up/down keys.
• Start the evaporation process by pressing the right-hand function
key "enter".
The start of the evaporation process starts the following automatic
process run:
• The rotor speed increases (direction of rotation: anti-clockwise)
• At 70% of the selected maximum speed: the aeration valve closes
automatically (to avoid boiling retardation).
• At 80% of the selected maximum speed: the stop valve opens the
connection to the vacuum pump
• Further speed increase until the preselected maximum rotor speed
is reached.
7. Interrupting an evaporation process:
• Press the left-hand function key "mode".
• Select "stop evaporation" and confirm with the right-hand function
key.
Without an interruption, the evaporation process ends after the
preselected time. The following process runs automatically:
• At maximum speed: the stop valve closes the connection to the
vacuum pump, the aeration valve opens; the rotor speed is
maintained for 30 sec (aeration phase, to avoid boiling retardation).
• The rotor decelerates until it is at a standstill. After the standstill of
the rotor the lid can be opened in order to remove the rotors.