Rotational Vacuum Concentrator RVC 2-18 CDplus
HCl resistant version
6 Operation
48 / 84
Version 04/2012, Rev. 2.3 of 03/07/2017
• sb
Translation of the original operating manual
Settings
The Settings menu is used to customise the operation and process
management of the control system.
• Select the desired menu.
• Change the value using the up/down keys.
• Confirm the input by pressing the right-hand function key.
Fig. 34: Settings menu
Timer warm-up
You can set a warm-up time between 5 and 99 minutes.
Temperature warm-up
During the warm-up phase, the rotor temperature is set to the set
temperature.
Click at button push
If this function is activated, a brief sound signal can be heard whenever a
key is pressed.
High temperature resolution
Temperatures are displayed in the values window with a resolution of
1
/
10
°C
(instead of a resolution of 1°C).
Rotor speed
The rotor speed can be set in a range between 100 rpm and 1,750 rpm.
NOTE
We guarantee that a speed of 1,350 rpm can be reached with all of the
available rotors. Higher speeds are possible in individual cases.
Continuous operation VP
If this option is
set to “yes”, the vacuum pump will not be switched off after
the evaporation. Instead, it will remain active. This is useful if several
evaporation processes are started in succession.