Later installation of a vacuum sensor
In freeze-dryers that have been delivered without a vacuum sensor, the
device option "Sensor: vacuum not installed" is activated at the factory in
order to suppress error messages during operation.
If a vacuum sensor is installed at a later time, this option must be
deactivated as described hereinabove.
Fig. 42: Device option "Sensor: vacuum not installed"
Service
"Service" options:
1
"Operating hours"
2
"Failure memory"
3
"Vacuum pump oil
change"
4
"Maintenance"
5
"Sensor adjustment"
Fig. 43: Dialogue box "Service" (varies depending on the type of freeze-dryer)
Operating hours (1)
This dialogue box is used to view the number of operating hours of the
various components of the freeze-dryer, e.g. the refrigeration unit, vacuum
pump or pressure control valve. In addition to the name, the respective
→
reference designation
is also displayed.
These data are provided for informational purposes only. They cannot be
edited.