4.1 Replacing the soda lime cartridge
When the bluing of the soda lime reaches the mark, this means it is
saturated in CO
2
, and must be replaced. On average, one cartridge
is good for approximately 50 tests
z
Procedure:
Open the cartridge lid
Then press on the two (shown in red)
connection clips by pulling on the
cartridge.
In the same way as you have taken
off the cartridge, put it back by
pressing on the clips.
Check that if you are opposite the
Rheo F4, the cartridge metal tube is
positioned as shown opposite.
Close the cartridge lid.
Once the cartridge has been filled, you must check the
tightness of the pneumatic circuit by following the
instructions in the "Leak test" chapter, page
35
.