background image

Fingertip  

Pulse Oximeter

 

USER MANUAL                  

Ver2.0C15D-H 

General Description 

Oxygen Saturation is a percentage of Oxyhemoglobin (HbO

2

) capacity, compounded with oxygen, by all combinative hemoglobin 

(Hb) capacity in blood. In other words, it is consistency of Oxyhemoglobin in blood. It is a very important parameter for the 
Respiratory Circulation System. Many respiratory diseases can result in oxygen saturation being lowered in human blood. 
Additionally, the following factors can reduce oxygen saturation: Automatic regulation of organ dysfunction caused by Anesthesia, 
Intensive Postoperative Trauma, injuries caused by some medical examinations. That situation might result in light-headedness, 
asthenia, and vomiting. Therefore, it is very important to know the oxygen saturation of a patient so that doctors can find problems 
in a timely manner. 

The fingertip pulse oximeter features low power consumption, convenient operation and portability. Place one fingertip into the 
photoelectric sensor for diagnosis and the pulse rate and oxygen saturation will appear on the display. It has been proven in clinical 
experiments that it also features high precision and repeatability. 

Measurement Principle 

Principle of the oximeter is as follows: A mathematical formula is established making use of Lambert Beer Law according to 
Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO

2

) in red and near-infrared zones. 

Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity 
Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm red and 905nm near 
infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. A measured signal obtained by a 
photosensitive element, will be shown on the oximeter’s display through process in electronic circuits and microprocessor. 

Diagram of Operation Principle 

1. Red and Infrared-ray Emission Tube 
2. Red and Infrared-ray Receipt Tube 

Precautions For Use 

Before use, carefully read the manual. 

Operation of the fingertip pulse oximeter may be affected by the use of an electrosurgical unit (ESU). 

The fingertip pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO

2

 measurement. Verify 

that nothing is hindering the pulse measurement before relying on the SpO

2

 measurement. 

Do not use the fingertip pulse oximeter in an MRI or CT environment. 

Do not use the fingertip pulse oximeter in situations where alarms are required. The device has no alarms. It is not for 
continuous monitoring. 

Do not use the fingertip pulse oximeter in an explosive atmosphere. 

The fingertip pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other 
methods of assessing clinical signs and symptoms. 

In order to ensure correct sensor alignment and skin integrity, the maximum application time at a single site for our device 
should be less than 4 fours. 

Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device in liquid. The device is not 
intended for sterilization. 

10 

Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components, 
including batteries. 

11 

This equipment complies with IEC 60601-1-2:2007 for electromagnetic compatibility for medical electrical equipment and/or 
systems. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical 
noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or 
strength of a source might disrupt the performance of this device. 

12 

Portable and mobile RF communications equipment can affect medical electrical equipment. 

13 

This equipment should not be used adjacent to or stacked with other equiopment. 

14 

This equipment is not intended for use during patient transport outside the healthcare facility. 

15 

Do not disassemble, repair or modify the equipment without authority. 

16 

These materials that contact with the patient’s skin contain medical silicone and ABS plastic enclosure are all pass the 
ISO10993-5 Tests for invitro cytotoxicity and ISO10993-10 Tests for irritation and delayed-type hypersensitivity. 

Rx only: “Caution: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.” 

Inaccurate measurements may be caused by 

Significant levels of dysfunctional hemoglobin (such as carbonyl - hemoglobin or methemoglobin). 

Intravascular dyes such as indocyanine green or methylene blue. 

High ambient light. Shield the sensor area if necessary. 

Excessive patient movement. 

High-frequency electrosurgical interference and defibrillators. 

Venous pulsations. 

Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line. 

The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia. 

The patient is in cardiac arrest or is in shock. 

10 

Fingernail polish or false fingernails. 

11 

Weak pulse quality (low perfusion). 

12 

Low hemoglobin. 

Product Features 

Simple to operate and convenient to carry. 

Small volume, light weight and low power consumption. 

LCD screen display SpO

2

, PR, and Pulse bar. 

2pcs AAA-size alkaline batteries; battery-low indicator. 

When no or low signal is detected, the pulse oximeter will power off automatically in 8 seconds. 

Intended Use 

Fingertip pulse oximeter is a portable non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin 
(SpO

2

) and pulse rate of adult and pediatric patients in hospitals, hospital-type facilities, and home environments. It is not for 

continuous monitoring. 

Operation Instructions 

Install two AAA batteries according to the Battery Installation instructions. 

Place one of your fingers into the rubber opening of the pulse oximeter. 

Press the switch button one time on front panel to turn the pulse oximeter on. 

Keep your hands still for the reading. Do not shake your finger during the test. It is recommended that you do not move your 
body while taking a reading. 

Read the data from the display screen. 

Product Accessories 

1. One 

lanyard 

2. Two 

AAA 

batteries 

3. 

One user’s manual

 

 

Front Panel 

The height of the bar graph indicates of the pulse and signal strength. The bar should be greater than 30% for a proper reading.  

 

Battery Installation 

1. 

Install two AAA batteries into the battery compartment. Match the plus (+) and minus (-) 
signs in the compartment. If the polarities are not matched, damage may be caused to 
the oximeter. 

2. 

Slide the battery door cover horizontally along the arrow shown as the picture. 

Notes:  

Please remove the batteries if the pulse oximeter will not be used for long periods of time. 

Using the Lanyard 

1. 

Thread thinner end of the lanyard through the loop. 

2. 

Thread thicker end of the lanyard through the threaded end before pulling it 
tightly. 

Warnings! 

 

Keep the oximeter away from young children. Small items such as the 
battery door, battery, and lanyard are choking hazards. 

 

Do not hang the lanyard from the device’s electrical wire. 

Maintenance and Storage 

1. 

Replace the batteries in a timely manner when low voltage lamp is lighted. 

2. 

Clean surface of the fingertip oximeter before it is used in diagnosis for patients. 

3. 

Remove the batteries if the oximeter is not operated for a long time. 

4. 

It is best to store the product in -20

℃~+

55

 

and 

93% humidity

.

 

5. 

Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage. 

6. 

Dispose of battery properly; follow any applicable local battery disposal laws. 

Cleaning the fingertip pulse oximeter 

Please use medical alcohol to clean the silicone touching the finger inside of oximeter with a soft cloth dampened with 70% 
isopropyl alcohol. Also clean the being tested finger using alcohol before and after each test. 

Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in the device. Allow the oximeter to 
dry thoroughly before reuse. 

The fingertip pulse oximeter requires no routine calibration or maintenance other than replacement of batteries. 

The use life of the device is five years when it is used for 15 measurements every day and 10 minutes per one 
measurement. Stop using and contact local service center if one of the following cases occurs: 

 

An error in the 

Possible Problems and solutions 

is displayed on screen. 

 

The oximeter cannot be powered on in any case and not the reasons of battery. 

 

There is a crack on the oximeter or damage on the display resulting readings cannot be identified; the spring is invalid; or the 

key is unresponsive or unavailable. 

A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor. Clinical testing is used to establish 
the SpO

2

 accuracy. The measured arterial hemoglobin saturation value (

SpO

2

) of the sensors is compared to arterial hemoglobin 

oxygen (

SaO

2

) value, determined from blood samples with a laboratory CO-oximeter. The accuracy of the sensors in comparison to 

the CO-oximeter samples measured over the 

SpO

2

 range of 70 

 100%. Accuracy data is calculated using the 

root-mean-squared (Arms value) for all subjects, per ISO 9919:2005, Medical Electrical Equipment

Particular requirements for the 

basic safety and essential performance of pulse oximeter equipment for medical use. 

A functional tester is used to measure how accurately Fingertip Pulse Oximeter is reproducing the specified calibration curve and 
the PR accuracy. 
The model of functional tester is Index2 FLUKE simulator and the version is 2.1.3. 

Specifications 

1. Display Type 

LCD display 

2. SpO

2

 

Display range: 35%

100% 

Measurement range: 70

100%  

Accuracy: 70-100%

±

2digits; 0%

69% no definition 

Resolution: 1%   

3. Pulse Rate 

Display range: 30

250BPM 

Measure range: 30

250 BPM 

Accuracy: 30

99bpm, ±2bpm;    100

250bpm, ±2% 

Resolution: 1BPM 

4. Probe LED Specifications 

 

Wavelength Radiant 

Power 

RED 660

±

2nm 1.8mW 

IR 905

±

10nm 2.0mW 

NOTE:

 The information about wavelength range can be especially useful to clinicians. 

5. Power Requirements 

Two AAA alkaline Batteries 
Power consumption: Less than 25mA   

Power indication:

 

 

Battery Life: Two AAA 1.5V, 800mAh alkaline batteries could be continuously operated as long as 24 hours.   

6. Environment Requirements 

Operation Temperature: 5

40

 

Storage Temperature: -20

~+

55

 

Ambient Humidity: 

80% no condensation in operation 

93% no condensation in storage 

Atmosphere pressure: 86kPa

106kPa 

7. Equipment Response Time 

MD300C15D-H 

Отзывы: