8
■
For accurate measurement
For accurate measurement, refer to the figure shown below and make sure that a measuring object
is imaged at the center of the display as large as possible. Larger measuring area needs longer
distance.
Ensure a sufficient measuring area for accurate measurement.
For accurate measurement, 4 x 4 pixels or more are needed for a target area.
■
Setting of emissivity
The emissivity setting can be changed by 3 digits (Nos. 6, 7 and 8 in the figure shown above) of the
8-digit slide switch placed on the bottom edge portion of the battery housing. The factory default
emissivity is 0.95.
5 kinds of the emissivity can be set from 0.80 to 1.00 with 0.05 increments.
Set the emissivity referring to the setting mode of the slide switch shown on the battery cover or the
table sown below.
No. 6
No. 7
No. 8
Emissivity
OFF
OFF
OFF
ε
=0.95
OFF
OFF
ON
ε
=0.90
OFF
ON
OFF
ε
=0.85
OFF
ON
ON
ε
=0.80
ON
OFF
OFF
ε
=1.00
About 12mm
Measuring view angle
About 37mm
■
33.3
°×
33.3
°
■
D (Viewing width) =
0.6
×
L (Measuring distance)
Field of view
Example of emissivity setting
8-digit slide switch