INST No. INE-412-P7CE
4.1 Cautions on installation
For installation, use holes indicated by (9) or (10).
4.2 Measuring distance and measuring diameter
4.3 Targeting
Target this thermometer to the measurement object by utilizing a
(1)
V-shaped groove on the upper side panel and
a
(3)
Sight line on the side panel.
4. Installation
Refer to [ 3. Each part name and outside dimensions]
Avoid installation in the following places.
・
Places receiving vibration or impact
・
Places where the ambient temperature is not 0 to 50ºC
・
Places having noises
・
Places where dusts, fine particles and other materials are surrounding
・
Places where this thermometer and the measurement objects receive heat radiation by
flames, sunlight and other sources
・
Places where there is dangerous gas such as combustible gas and explosion gas
Place this thermometer at the distance where the diameter of the measurement object is more
than 1.5 times of the measurement diameter calculated by the following relation.
When the temperature of the measurement object is higher than the ambient temperature,
change the direction of this thermometer slightly to the direction having the maximum
output. On this direction, you will have the accurate measurement.
C au t i on
C au t i on
Reference
Ø
9.5
Ø
5
Ø
7.5
Ø
15
60
80
100
I N F R A R E D T H E R M O M E T E R
I R- B
Ø
15
Ø
40
Ø
95
1000
500
I N F R A R E D T HE R M O M E T E R
I R- B
I N F R A R E D T HE R M O M E T E R
I R- B
Ø
15
Ø
20
Ø
40
Ø
60
Ø
80
500
1000
1500
2000
•IR-BAT1A, IR-BAT1B
(General: General type)
•IR-BAT1M, IR-BAT1P
(High speed: General type)
•IR-BAT2A
(General: Small diameter, short distance)
•IR-BAT2M
(High speed small diameter, short distance)
•IR-BATSA, IR-BATSB
(General: Small diameter, long distance)
•IR-BATSM, IR-BATSP
(High speed small diameter, long distance)
Unit: mm