Chino IR-BAT Series Скачать руководство пользователя страница 2

INST No. INE-412-P7CE

For the purpose of using this thermometer safely, observe the following cautions for safety
during the operation, maintenance, and repair of this thermometer.
Take safety measures separately against wrong operation, troubles, and other circumstances
of this thermometer if necessary.

This instruction manual uses the following symbol marks for safe.

CAUTIONS FOR SAFETY

Item

Cautions

Power supply

Make sure that the voltage of the power source meets the rated power supply of
this thermometer before operation.

Prohibition of use in a
gaseous atmosphere

Never operate this thermometer at a place where a combustible gas, an
explosive gas or their vapors exist, otherwise a very dangerous accident
will occur. It is extremely dangerous to use this thermometer under such
environments.

Never touch the interior

To prevent from trouble, don't touch the inside of this thermometer by hand.

Caution on external
connections

For connection to external equipment, make sure to turn off the power supply
and also read instruction manual for the equipments.

Prohibition disassembling
and modification

Don't disassemble nor modify this thermometer. If you disassemble or
modify this thermometer, it will be dangerous to causing the trouble.

Cautions on maintenance
and check of the lens

When wiping out the dirt of the lens, don't use organics solvents except an
alcohol, water and the detergents absolutely.
If used, it will cause the trouble.

Type

Model

Measuring
range

Response
time (*)

Diameter/

Distance

Measuring
wavelength

G

en

er

a

l

General type

IR-BAT1A

0 to 300 ºC

100ms

Ø40/500mm

8 to 14

m

IR-BAT1B

0 to 600 ºC

Small diameter, short distance

IR-BAT2A

0 to 300 ºC

100ms

Ø5/80mm

Small diameter, long distance

IR-BATSA

0 to 300 ºC

100ms

Ø40/1000mm

IR-BATSB

0 to 600 ºC

Type

Model

Measuring
range

Response
time (*)

Measuring
diameter

Measuring
wavelength

H

ig

h

sp

ee

d

General type

IR-BAT1M

0 to 300 ºC

50ms

Ø40/500mm

8 to 14

m

IR-BAT1P

0 to 600 ºC

Small diameter, short distance IR-BAT2M

0 to 300 ºC

50ms

Ø5/80mm

Small diameter, long distance

IR-BATSM

0 to 300 ºC

50ms

Ø40/1000mm

IR-BATSP

0 to 600 ºC

Observe this caution item strictly for avoiding a danger, otherwise negligence of this
caution may cause an electric shock accident or other accidents resulting in
operator’s death or injuries.
Observe this caution item strictly, otherwise negligence of this caution may cause the
damage of this thermometer or measuring errors.
This mark is attached to each title covering

instructions

This mark shows the act of the prohibition.

W a r n i ng

C au t i on

Warnin g

FOR SAFE USE

1.Models and main specifications

(*Response time: 95

)

Содержание IR-BAT Series

Страница 1: ... necessary for maintenance too Keep this manual with due care until this thermometer is discarded NOTE 1 The descriptions of this manual are subject to change without notice 2 If a question has arisen or if an omission was found in this manual please contact your nearest CHINO s sales agent 3 We are not responsible for any results by operation of this thermometer Thank you for your purchase of IR ...

Страница 2: ...l be dangerous to causing the trouble Cautions on maintenance and check of the lens When wiping out the dirt of the lens don t use organics solvents except an alcohol water and the detergents absolutely If used it will cause the trouble Type Model Measuring range Response time Diameter Distance Measuring wavelength General General type IR BAT1A 0 to 300 ºC 100ms Ø40 500mm 8 to 14 m IR BAT1B 0 to 6...

Страница 3: ...h part name and outside dimensions 6 7 Control panel 15 85 f 26 I N F R A R E D T H E R M O M E T E R I R B 21 1 55 32 22 2 3 4 5 1 25 1 13 10 11 12 4 M2 Screw for control panel cover 8 2 M4Depth 4 9 1 4w Depth 4 BA019T 004 IR BAT1A 0 300 03 BA044T 004 IR BAT1A 0 300 03 Temperature range Software version Model Serial No Serial No Label CE mark 1 V shaped groove 2 Lens 3 Sight line 4 3mm O tip 5 Ca...

Страница 4: ...s and explosion gas Place this thermometer at the distance where the diameter of the measurement object is more than 1 5 times of the measurement diameter calculated by the following relation When the temperature of the measurement object is higher than the ambient temperature change the direction of this thermometer slightly to the direction having the maximum output On this direction you will ha...

Страница 5: ... IR GBHA1 Indicator with power supply and IR GC1 Indicator with power supply and Model IR GBG2 Converter with reflection compensation refer to the instruction manual for each model Reference 6 Operation Caution If the temperature of the measurement object is unknown measure it with a thermocouple and adjust Emissivity setting digital switches Adjustments may not be possible in the following cases ...

Страница 6: ...e heat sources 5 Remove the dirt from the lens 6 Adjust the target 7 Check and maintenance Caution When wiping out the dirt of the lens don t use organic solvents except alcohol water and detergent absolutely If used it will cause the trouble 8 Trouble shooting 9 Cautions Caution 1 Install this thermometer at places where the ambient temperature is 0 to 50ºC 2 Avoid to sets this thermometer at pla...

Страница 7: ...0 Inner diameter Ø4 Connection cable inlet Purge air inlet Outer diameter Ø6 L metal fittings Ø 4 3 10 5 52 24 125 33 45 Ø 29 40 39 113 2 hole for mounting L angle After opening case fix it To the thermometer top 2 57 27 80 10 2 Protective case Thermometer by M4 screw Inner diameter Ø4 Purge air inlet Outer diameter Ø6 Ø 4 3 10 5 Connection cable inlet L metal fittings 2 hole for mounting L angle ...

Страница 8: ...2000mm IR ZBRA Max length is 200m Out of CE Marking Case Aluminum die cast Waterproof JIS C 0920 Water drop protection II type IP X2 Mounting method By 2 M4 screws or on tripod Weight About 220g CE EN61326 1 Class A Conditions for CE conformance 1 The connection cable should be used indoors and its length should be up to 30m 2 One 1 set of DC power source unit should be individually connected to o...

Отзывы: