Chino IR-BAT Series Скачать руководство пользователя страница 1

INST No. INE-412-P7CE

IR-BAT



series

COMPACT INFRARED
RADIATION THERMOMETER

Model:IR-BAT1A, IR-BAT1B, IR-BAT2A

IR-BATSA, IR-BATSB
IR-BAT1M, IR-BAT1P, IR-BAT2M
IR-BATSM, IR-BATSP

PREFACE/

Request and notices

Request to designers, instrument controllers, and sale agents

Make sure to deliver this instruction manual to the operator of this thermometer

Request to the operator of this instrument

This instruction manual is necessary for maintenance, too.

Keep this manual with due care until this thermometer is discarded.

NOTE

1.The descriptions of this manual are subject to change without

notice.

2.If a question has arisen or if an omission was found in this

manual, please contact your nearest CHINO’s sales agent.

3.We are not responsible for any results by operation of this

thermometer.

Thank you for your purchase of IR-BAT



series Compact

infrared radiation thermometer.
Please read this instruction manual for using this thermometer
correctly, safely and also preventing troubles in advance.

marking

EMC directive.
EN61326

-

1 Class A

Содержание IR-BAT Series

Страница 1: ... necessary for maintenance too Keep this manual with due care until this thermometer is discarded NOTE 1 The descriptions of this manual are subject to change without notice 2 If a question has arisen or if an omission was found in this manual please contact your nearest CHINO s sales agent 3 We are not responsible for any results by operation of this thermometer Thank you for your purchase of IR ...

Страница 2: ...l be dangerous to causing the trouble Cautions on maintenance and check of the lens When wiping out the dirt of the lens don t use organics solvents except an alcohol water and the detergents absolutely If used it will cause the trouble Type Model Measuring range Response time Diameter Distance Measuring wavelength General General type IR BAT1A 0 to 300 ºC 100ms Ø40 500mm 8 to 14 m IR BAT1B 0 to 6...

Страница 3: ...h part name and outside dimensions 6 7 Control panel 15 85 f 26 I N F R A R E D T H E R M O M E T E R I R B 21 1 55 32 22 2 3 4 5 1 25 1 13 10 11 12 4 M2 Screw for control panel cover 8 2 M4Depth 4 9 1 4w Depth 4 BA019T 004 IR BAT1A 0 300 03 BA044T 004 IR BAT1A 0 300 03 Temperature range Software version Model Serial No Serial No Label CE mark 1 V shaped groove 2 Lens 3 Sight line 4 3mm O tip 5 Ca...

Страница 4: ...s and explosion gas Place this thermometer at the distance where the diameter of the measurement object is more than 1 5 times of the measurement diameter calculated by the following relation When the temperature of the measurement object is higher than the ambient temperature change the direction of this thermometer slightly to the direction having the maximum output On this direction you will ha...

Страница 5: ... IR GBHA1 Indicator with power supply and IR GC1 Indicator with power supply and Model IR GBG2 Converter with reflection compensation refer to the instruction manual for each model Reference 6 Operation Caution If the temperature of the measurement object is unknown measure it with a thermocouple and adjust Emissivity setting digital switches Adjustments may not be possible in the following cases ...

Страница 6: ...e heat sources 5 Remove the dirt from the lens 6 Adjust the target 7 Check and maintenance Caution When wiping out the dirt of the lens don t use organic solvents except alcohol water and detergent absolutely If used it will cause the trouble 8 Trouble shooting 9 Cautions Caution 1 Install this thermometer at places where the ambient temperature is 0 to 50ºC 2 Avoid to sets this thermometer at pla...

Страница 7: ...0 Inner diameter Ø4 Connection cable inlet Purge air inlet Outer diameter Ø6 L metal fittings Ø 4 3 10 5 52 24 125 33 45 Ø 29 40 39 113 2 hole for mounting L angle After opening case fix it To the thermometer top 2 57 27 80 10 2 Protective case Thermometer by M4 screw Inner diameter Ø4 Purge air inlet Outer diameter Ø6 Ø 4 3 10 5 Connection cable inlet L metal fittings 2 hole for mounting L angle ...

Страница 8: ...2000mm IR ZBRA Max length is 200m Out of CE Marking Case Aluminum die cast Waterproof JIS C 0920 Water drop protection II type IP X2 Mounting method By 2 M4 screws or on tripod Weight About 220g CE EN61326 1 Class A Conditions for CE conformance 1 The connection cable should be used indoors and its length should be up to 30m 2 One 1 set of DC power source unit should be individually connected to o...

Отзывы: