6 Measurement Optimization
6.1 Reduction of Accessory Influence
183
6 Measurement Optimization
The following methods can be applied to adjust the setting and optimize the measurement accuracy.
●
Reduction of accessory influence
·················································································
●
Improvement of reflection accuracy of low-loss 2-port device
·················································
●
Increase of dynamic range
························································································
●
Improvement of measurement results of electrical long device
················································
●
Improvement of phase measurement accuracy
··································································
●
Reduction of trace noise
···························································································
●
Increase of data point number
·····················································································
●
Improvement of measurement stability
···········································································
●
Improvement of sweep velocity
···················································································
●
Improvement of efficiency of multi-state measurement
·························································
●
Rapid data transmission
···························································································
●
Macro use
6.1 Reduction of Accessory Influence
Some accessories may affect the measurement results of the DUT. Select one of two characteristics of the analyzer
to reduce the accessory influence. In order to reduce the accessory influence, you can select the most suitable
method to test the configuration.
1) Compensate the cable loss with the power slope.
If a long cable or other kind of accessory is used in measurement configuration, the power loss will be generated
within the whole band. In this case, the power slope function should be applied. The defined ratio (dB/GHz) can be
used in this function to increase the source power of the analyzer. The process of cable loss compensation is as
follows.
Menu path:
[Stimulus] > [Power...]
.
If the power slope function is not activated, tick the
[Power Slope]
check box. Input the ratio to be increased in
frequency sweep of the source power in the
[dB/GHz]
box.
2) Selectively remove the response gate (only for the time domain).
a) The gate is one of the time domain characteristics. It is used to remove the response mathematically. The gate of
reflection response or transmission response can be set, but different results will be obtained.
➢
Gate reflection response
: the desired response (such as the return loss of the filter) can be separated
from the undesired response (such as the adapter reflection or connector mismatch) by reflection response
of the gate.
➢
Gate transmission response
: the specific path of large electrical length in the multi-path device can be
separated by transmission response of the gate.
In order to apply the gate in the time domain mode, the transformation can be disconnected, and the existing gate
pulse can be used in study of the frequency response of the device. The time filter gate pulse has the following
shapes and features:
➢
Band pass shape
: the response beyond the passband zone can be removed.
➢
Trap shape
: unnecessary response in the frequency zone can be removed.
b) Removal of unnecessary response
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...