![China Electronics Technology Instruments AV3672 Series Скачать руководство пользователя страница 203](http://html1.mh-extra.com/html/china-electronics-technology-instruments/av3672-series/av3672-series_user-manual_2601790203.webp)
6 Measurement Optimization
6.8 Improvement of Measurement Stability
199
Fig. 6.18 Setting of Number of Sweep Points
2) Optimization of measurement results
One measurement includes several interdependent settings. You can modify the settings to achieve the purposes of
measurement application: faster throughput or more accurate data. You can balance the settings to optimize
measurement.
➢
Increase the measurement throughput.
➢
Improve the measurement accuracy.
6.8 Improvement of Measurement Stability
Measurement instability is caused under several conditions. For repeated measurement, you can apply several
methods to create a stable measurement environment. The following factors have adverse effects on the
measurement accuracy.
1) Frequency offset
The frequency precision of the analyzer depends on the accuracy of the internal 10MHZ oscillator. If high
frequency accuracy and stability are required in the measurement application, no consideration should be given to
the internal frequency standard, and a high-stability external frequency source can be created by the 10MHz
reference input connector on the rear panel.
2) Temperature deviation
a) The electrical characteristics of the following components can be changed as a result of thermal expansion and
contraction.
➢
Internal device of analyzer
➢
Standard of calibration kit
➢
Tester
➢
Cable
➢
Adapter
b) The temperature deviation in measurement can be reduced in the following methods.
➢
Apply a temperature-controlled environment.
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...