4 Measurement Setting
4.6 Trigger Mode
98
Fig. 4.15Sweep Setting
2) [Sweep Setup] dialog box
a) [Channel] box
Select the channel applicable to sweep setting.
b) [Stepped Sweep] check box
In the
step sweep
mode, the source is tuned to a frequency point and the frequency point is measured after the
specified dwell time. Then the source is tuned to next frequency point. A device of long electrical delay can be
accurately measured. If this check box is cleared, the analyzer may work in the
analog sweep
or
step sweep
mode,
depending on the setting of the sweep time and intermediate frequency bandwidth.
c) [Dwell time] input box
Set the dwell time at each point before obtaining the measurement data. This is only applicable to the
step sweep
mode.
d) [Sweep Delay] input box
Set the waiting time of the analyzer before obtaining the measurement data. It refers to the delay before
measurement of the first point.
e) [Fast Sweep] check box
If the
[Fast Sweep]
check box is selected, analog sweep will be started. Data are collected in the sweep process,
thus reducing the waiting time of sweep.
4.6 Trigger Mode
The trigger signal is used to enable measurement sweep of the analyzer. The sweep mode and time to stop sweep
and return to the holding state are determined by trigger settings. The vector network analyzer has high flexibility
in trigger setting.
●
Simple trigger setting
································································································
●
Detailed trigger setting
······························································································
●
[Trigger] dialog box
·······························································································
4.6.1 Simple Trigger Setting
Only the trigger mode of the current active channel can be set by simple trigger setting.
Menu path:
[Stimulus] > [Trigger]
. The
trigger
sub-menu will appear.
Click
[Continuous], [Single] or [Hold]
in the sub-menu and select the trigger mode.
Содержание AV3672 Series
Страница 1: ...AV3672 Series Vector Network Analyzer User Manual China Electronics Technology Instruments Co Ltd...
Страница 3: ......
Страница 4: ...AV3672 Series Vector Network Analyzer Contents...
Страница 5: ......
Страница 124: ...5 Menu 5 1 Menu structure 120 5 1 2 Track Fig 5 2 Track Menu...
Страница 125: ...5 Menu 5 1 Menu structure 121 5 1 3 Channel Fig 5 3 Channel Menu...
Страница 126: ...5 Menu 5 1 Menu structure 122 5 1 4 Excitation Fig 5 4 Excitation Menu I...
Страница 127: ...5 Menu 5 1 Menu structure 123 Fig 5 5 Excitation Menu II...
Страница 128: ...5 Menu 5 1 Menu structure 124 Fig 5 6 Excitation Menu III...
Страница 129: ...5 Menu 5 1 Menu structure 125 5 1 5 Response Fig 5 7 Response Menu I...
Страница 130: ...5 Menu 5 1 Menu structure 126 Fig 5 8 Repsonse Menu II...
Страница 131: ...5 Menu 5 1 Menu structure 127 Fig 5 9 Response Menu III...
Страница 132: ...5 Menu 5 1 Menu structure 128 Fig 5 10 Response Menu V Fig 5 11 Response IV...
Страница 133: ...5 Menu 5 1 Menu structure 129 5 1 6 Calibration Fig 5 12 Calibration Menu...
Страница 134: ...5 Menu 5 1 Menu structure 130 5 1 7 Marker Fig 5 13 Cursor Menu I...
Страница 135: ...5 Menu 5 1 Menu structure 131 Fig 5 13 Cursor Menu II...
Страница 136: ...5 Menu 5 1 Menu structure 132 Fig 5 15Marker Menu III...
Страница 137: ...5 Menu 5 1 Menu structure 133 5 1 8 Analysis Fig 5 16 Analysis Menu I...
Страница 138: ...5 Menu 5 1 Menu structure 134 Fig 5 17 Analysis Menu II...
Страница 139: ...5 Menu 5 1 Menu structure 135 Fig 5 18 Analysis Menu III...
Страница 140: ...5 Menu 5 1 Menu structure 136 5 1 9 System Fig 5 19 System Menu I...
Страница 141: ...5 Menu 5 1 Menu structure 137 Fig 5 20 System Menu I...
Страница 254: ...8 Basis of Network Measurement 8 3 Amplifier Parameter Specifications 250...
Страница 257: ...8 Basis of Network Measurement 8 4 Complex Impedance 253...
Страница 366: ...Appendix Appendix 3 Frequency Offset Measurement 362 Fig 3 13 Test of Return Loss of Mixer LO Port...
Страница 373: ...Appendix Appendix 4 Pulse Measurement 369 Fig 4 9 Receiver gain configuration Dialog Box...